Parameter Settings for the Optional Interface Port 

o  Factory Settings

5.  Interface 

5.1

Baud rate

5.1.2

300 baud

5.1.3

600 baud

5.1.4

1200 baud

5.1.5

o

2400 baud

5.1.6

4800 baud

5.1.7

9600 baud

5.1.8

19200 baud

5.2

Parity

5.2.3

o

Odd 

5.2.4

Even 

5.2.5

None 

5.3

Number of stop bits

5.3.1

o

1 stop bit

5.3.2

2 stop bits

5.4

Handshake 5.4.1

Software 

handshake

mode

5.4.2

o

Hardware handshake

5.4.3

None

6.Printing

6.1 Data output 

6.1.1

On request regardless of stability

6.1.2

On request after stability

6.1.3

o

Auto print regardless of stability

6.1.4

Auto print after stability

Summary of Contents for PMA.Economy

Page 1: ...98648 016 10 Economy PMA Installation Instructions Sartorius PMA Economy Model PMA5000 PMA5000 000U Electronic Paint mixing Scales...

Page 2: ...ection to AC power 8 USB connection optional 9 Column with channels for power and USB cables 10 Weighing pan 11 Foot The following symbols are used in these instructions Indicates required steps Indic...

Page 3: ...0 PMA5000 000U into operation Warnings and Safety Precautions This scale meets the guidelines and stan dards governing electronic equipment electromagnetic compatibility and the stipulated safety stan...

Page 4: ...ousing must have foil shielding that is suitable for high fequen cies in accordance with CE regulations The operator shall be solely responsible for any damage or injuries that occur when using cables...

Page 5: ...pan AC adapter or power supply USB cable only supplied with the PMA5000 000U Setting Up the Scale Choose a suitable place to set up the scale Avoid exposure to drafts heat moisture and vibration Make...

Page 6: ...ns and warnings in this manual Check that the voltage rating of the AC adapter matches that of your local line voltage mains supply If it does not match your local voltage rating or if the plug design...

Page 7: ...your formula is reached Remove the filled paint can from the weighing pan Never use a hammer to close the lid of a paint can while it is still on the weighing pan Otherwise you will damage the weighin...

Page 8: ...y 6 Example calibration weight 5000 g Confirm the calibration weight selected by pressing the L key ENTER 5 After the scale has been tared place the displayed calibration weight on the weighing pan Ce...

Page 9: ...2 Application filter 1 2 1 Tare 1 2 2 o Dose 1 3 Stability range 1 3 1 1 4 increment step 1 3 2 1 2 increment step 1 3 3 1 increment step 1 3 4 o 2 increment steps 1 3 5 4 increment steps 1 5 Calibrat...

Page 10: ...0 baud 5 1 8 19200 baud 5 2 Parity 5 2 3 o Odd 5 2 4 Even 5 2 5 None 5 3 Number of stop bits 5 3 1 o 1 stop bit 5 3 2 2 stop bits 5 4 Handshake 5 4 1 Software handshake mode 5 4 2 o Hardware handshake...

Page 11: ...e not met e g Zero tare with the U key Zero tare Do not calibrate until a zero readout Weighing pan is loaded is displayed appErr Too little weight or in the Increase weight case of application progra...

Page 12: ...to the AC adapter power supply or cables The AC adapter power supply no longer functions properly The AC adapter power supply has been stored for a relatively long period under unfavorable conditions...

Page 13: ...t members of the European Economic Area EEA or where no Sartorius affiliates subsidiaries dealers or distributors are located please contact your local authorities or a com mercial disposal operator P...

Page 14: ...mes equipped with an integrated interface port Disconnect power supply Unplug the AC adapter Turn the scale onto its side Slide off the plate located on the back of the display and control unit Insert...

Page 15: ...n of Windows may have slightly different procedures for upload ing the driver from the CD All versions have a Wizard for adding and removing hardware which guides you through the process 3 After you c...

Page 16: ...kg 1 class M1 or better Power consumption VA Average 1 6 maximum 5 8 Interface port optional USB Format 7 bit ASCII 1 start bit 1 or 2 stop bits Parity Even odd or no parity Transmission rates 300 to...

Page 17: ...17 C Marking...

Page 18: ...ngen Germany All rights reserved No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius AG The status of the informati...

Reviews: