18

Care and Maintenance

Cleaning

!

Never use concentrated acids, alkali solu-
tions or pure alcohol to clean the equip-
ment.

!

Do not allow liquids to penetrate the
equipment housing.

$

Use a brush or a soft, dry, lint-free cloth
to clean the scale.

Cleaning Stainless Steel Surfaces

Clean all stainless steel parts regularly.
Remove the stainless steel load plate and
thoroughly clean it separately. Use a damp
cloth or sponge to clean stainless steel
parts on the scale. You can use any com-
mercially available household cleaning
agent that is suitable for use on stainless
steel. Clean stainless steel surfaces by
wiping them down. Then clean the load
plate thoroughly, making sure to remove
all residues. Afterwards, wipe down stain-
less steel parts again using a clean, damp
cloth or sponge and allow the equipment
to dry. If desired, you can apply oil to the
cleaned surfaces as additional protection.

!

Solvents are permitted only for cleaning
stainless steel parts. 

Corrosive Environment

$

Remove all traces of corrosive substances
from the scale on a regular basis. 

Storage and Shipping Conditions

$

The packaging used for shipping your
Sartorius equipment is optimally designed
to prevent damage during transport. 
It is a good idea to save the box and all
parts of the packaging for future storage
or shipment of the equipment. Only the
original packaging provides the best 
protection for shipment.

$

Allowable storage temperature: 
–20 °C to +75 °C (–4 °F to +167 °F)

$

Allowable humidity during storage: 
up to 90% 

$

Please refer to the information under
“Safety Inspection” below.

Safety Inspection

Safe operation of the scale is no longer
ensured in the following cases: 

– If there is visible damage to the connect-

ing cable

– If the device no longer functions properly
– If the equipment has been stored for a

relatively long period under unfavorable
conditions

– If the equipment has been subjected to

rough handling during shipment

§

Make sure all warnings and safety 
precautions are observed, and notify 
your nearest Sartorius Service Center 
or the International Technical Support
Unit based in Goettingen, Germany, 
if any of the above occurs. Maintenance
and repair work may be performed only
by authorized Sartorius service techni-
cians who have access to the required
maintenance manuals and have received
the necessary training.

!

Only authorized service technicians may
open the equipment and perform mainte-
nance work; this ensures that operation
of the equipment is safe and trouble-free
and the warranty remains in effect.

Summary of Contents for PMA.Power PMA35001

Page 1: ...98648 016 91 Installation Instructions Sartorius PMA Power Model PMA35001 Electronic Paint mixing Scales...

Page 2: ...applications depends on the menu setting 8 L ENTER key and MEM key for paint mixing applications 9 Mounting bracket for display and control unit 10 Interface 25 contact D Sub socket 11 Grounding termi...

Page 3: ...y designed for use in paint mixing applications This scale can be operated through the keypad in stand alone operation or using application software such as a paint mixing program from a paint manufac...

Page 4: ...d differently The operator shall be solely responsible for any damage or injuries that occur when using cables not supplied by Sartorius Connect only Sartorius accessories and options as these are opt...

Page 5: ...s at the place of installation To ensure that the equipment is placed in a perfectly horizontal position for consistently reproducible weighing results Always level the weighing platform again any tim...

Page 6: ...ng or stan dard you use please contact your Sartorius office or dealer Use only original Sartorius power cables Use of cables from other manufacturers even if these units have a registered approval ra...

Page 7: ...this example g is displayed Add the other components up to the desired weight formula Remove the filled paint can from the load plate Never use a hammer to close a paint can while is it still on the l...

Page 8: ...of 1 0 to 5 0 or in 0 01 steps for factors of up to 1 0 Note A flashing arrow n on the display indicates that the weight value displayed is not valid in legal metrology i e not legal for trade Example...

Page 9: ...e overpour occurred After the correction has been completed you can continue filling the remaining components Note You can correct overpours as often as needed Keep in mind that the total quantity of...

Page 10: ...1 The readout shows the amount of the first compo nent to be added and C1 is shown in the upper right hand corner of the display 1 7 g 12 Add the first component until the readout shows 0 0 g 13 Press...

Page 11: ...lways allow approximately 30 minutes for the scale to warm up after connecting it to AC power and before performing calibration adjustment Press the U key zero tare for 2 seconds 10000 is displayed Re...

Page 12: ...ect the desired menu item on this level Press the L key ENTER to select the next menu level The third menu level is displayed Use the RS keys to select the desired menu item on this level Press the L...

Page 13: ...Level 3 Level 4 SETUP Press the L key ENTER SCALE Press the L key ENTER UNIT Use the RS keys to select the desired setting e g UNIT o GRAMS Press the L key ENTER PT PD Use the RS keys to select the de...

Page 14: ...active setting Press the c key Clear Activating the LOCK Function The LOCK function protects the scale from unauthorized use When this function is active the scale readout shows weight values only wh...

Page 15: ...f your password in a safe place The LOCK function can be deactivated only with this password Level 1 Level 2 Level 3 INPUT Use the RS keys to select INPUT PASSWORD Press the L key ENTER Press the L ke...

Page 16: ...ord protected Level 1 Level 2 Level 3 Level 4 SETUP Press the L key ENTER RESET Use the RS keys to select REST Press the L key ENTER MENU Use the RS keys to select MENU Press the L key ENTER YES Use t...

Page 17: ...f changes constantly installation Too much vibration or the Access the menu to select scale is exposed to a draft the appropriate code for the ambient conditions see Oper ating Menu Configuration The...

Page 18: ...s optimally designed to prevent damage during transport It is a good idea to save the box and all parts of the packaging for future storage or shipment of the equipment Only the original packaging pro...

Page 19: ...e technicians on location or our Service Center in Goettingen Germany Sartorius Service Center Weender Landstrasse 94 108 37075 Goettingen Germany Prior to disposal and or scrapping of the equipment a...

Page 20: ...for operation of the scale with a PC If you wish to create your own application software Sartorius can supply the required drivers for DOS or Windows operating systems Make sure to observe the safety...

Page 21: ...243 132 5 Net weight approximate kg 11 4 Calibration weight kg 10 class E2 or better Power consumption VA Average 8 maximum 16 Interface port RS 232 Format 7 bit ASCII 1 start bit 1 or 2 stop bits Pa...

Page 22: ...22...

Page 23: ...23...

Page 24: ...Goettingen Germany All rights reserved No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius The status of the infor...

Reviews: