Operating Instructions PMA 35001-X 

27

 Specifications 

Specifications

Model

PMA 35001-X

Weighing Capacity

g

35000

Readability

g

0.1

Tare range (subtractive)

g

-35000

Linearity

g

±0,2

Stability range (configurable in menu)

Digit

0.25 to 4

Humidity class

F

non-condensing

Permissible ambient operating temperature range

°C

0 to +40

IP protection

IP 43, in accordance with EN60529/IEC60529

Load plate dimensions

mm

350 

+

 240

Scale housing (WxDxH)

mm

350 

+

 243 

+

 132,5

Net weight, approx.

kg

11.4

External calibration weight

kg

10 (accuracy class F2 or better)

Interface

RS232

Format

7-bit ASCII, 1 start bit, 1 or 2 stop bits

Parity

Even, odd, none

Transmission rates

1200 to 38,400 bps

Handshake

Software or hardware

Power consumption

VA

Average: 8, max. 16

Mains Connection

V

100 – 240  V  ~

Voltage frequency

Hz

50 – 60

Device ID (explosion protection) II 2G Ex ia IIC T4 Gb 
DEKRA 12ATEX0180 X The terminal/control panel is 
suitable for use in potentially explosive areas in 
accordance with Directive 94/9/EC

 
 
 
Zone1 (devices in category 2)

Ambient conditions

Environment

For indoor use only

Ambient temperature: Storage and shipping

-10 °C  …  +60 °C

Ambient temperature: Operation

0 °C  …  +40 °C

Highest relative humidity:

80 % for temperatures up to 31 °C, decreasing linearly up to 
50 % relative humidity for 40 °C

Safety of electrical equipment

as per EN 61010-1:2010  
Safety requirements for electrical equipment for 
measurement, control, and laboratory use — Part 1: 
General requirements

Electromagnetic compatibility 

 

 

Defined immunity to interference: 
Limitation of emissions:

as per EN 61326-1:2006 
Electrical equipment for measurement, control, and labora-
tory use — EMC requirements — Part 1: General requirements 
Industrial areas 
Class B: Suitable for use in residential areas and areas that 
are directly connected to a low voltage network

Summary of Contents for PMA.Power Series

Page 1: ...Operating Instructions Sartorius PMA Power Model PMA35001 X Electronic Paint mixing Scale for Use in Potentially Explosive Atmospheres 1000025334...

Page 2: ...factor key for paint mixing applications 7 c key Clear and REC key for paint mixing applications 8 L ENTER key and MEM key for paint mixing applications Item Description 9 Mounting bracket for display...

Page 3: ...nt Supplied 8 Installation 8 Operation 12 Applications 13 Calibration 16 Menu Setting 17 Error Messages 22 Care and Maintenance 23 Disposal 25 Interface Port 26 Specifications 27 Accessories 28 Docume...

Page 4: ...manual from our website www sartorius com Warning Danger Symbols Warning danger symbols used in these instructions This symbol identifies hazards which have a high probability of resulting in death or...

Page 5: ...Improper use or handling however can result in damage and or injury The manufacturer is not responsible for any damage caused by non compliance with warnings or safety instructions The requirements pe...

Page 6: ...e checked for correct function and safety by a trained and qualified person at appropriate intervals e g checking the cable for damage Operating personnel must be trained to recognize faulty operating...

Page 7: ...d personnel with the power disconnected The device is intended to be installed exclusively in locations that offer sufficient protection against the penetration of solid foreign bodies or water The sa...

Page 8: ...age If you detect any damage proceed as directed in the Safety Inspections chapter page 23 Installation The color mixing scale is authorized for use in Zone 1 potentially explosive areas see documents...

Page 9: ...acity and inductance values see appendix on EC Type Examination Certificate and the EMC behavior The system should only be operated for the first time when it is certain that the area is not potential...

Page 10: ...ice professionals should be contacted Leveling the Weighing Platform Purpose To compensate for unevenness at the place of installation To ensure that the equipment is placed in a perfectly horizontal...

Page 11: ...installation of devices in Zone 1 In order to connect the equipment to the power supply use a correctly installed wall outlet with a protective grounding conductor PE and a fuse of a maximum 16 A t Co...

Page 12: ...lace an empty paint can on the load plate Press the U key Zero Tare 3 The display shows 0 0 g Pour the first component of your formula into the can and read off the weight when the stability symbol in...

Page 13: ...or of up to 1 0 Note A flashing n arrow on the display indicates that the weight value displayed is not valid in legal metrology i e not legal for trade Example As you pour in the components of your f...

Page 14: ...lues you entered and saved by pressing the L key MEM Press the S key to start the recalculation program C flashes on the display Press the S key to correct the value so that it matches the amount spec...

Page 15: ...repeatedly to correct the value 200 0 g 10 Press the L key MEM COR 01 11 Add the required amount of the first component C1 appears on the display 1 7 g 12 Add until value reaches 0 0 g 0 0 g 13 Press...

Page 16: ...precision 2 Always allow approximately 30 minutes for the scale to warm up after connecting it to AC power and before performing calibration adjustment Press and hold the U key Zero Tare for 2 second...

Page 17: ...n level two Use the RS keys to select the desired menu item t Use the L key ENTER to select the next menu level level three t The third level menu is displayed Use the RS keys to select the desired me...

Page 18: ...etting SETUP Press the L key ENTER SCALE Press the L key ENTER UNIT Use the RS keys to select e g SET GRAMS Press the L key ENTER KG o PT PD Use the RS keys toselect e g STANDARD SET Press the L key E...

Page 19: ...key ENTER o indicates the active setting Press the c key Clear Activating the LOCK Function The LOCK function protects the scale from unauthorized use When this function is active the scale readout sh...

Page 20: ...Keep a copy of your password in a safe place The LOCK function can be deactivated only with this password Level 1 Level 2 Level 3 Setting INPUT Use the RS keys to select INPUT PASSWORD Press the L key...

Page 21: ...feature is password protected Level 1 Level 2 Level 3 Level 4 Setting SETUP Press the L key ENTER RESET Use the RS keys to select RESET Press the L key ENTER MENU Use the RS keys to select MENU Press...

Page 22: ...hanges constantly The setup location is unstable Excessive vibration or draft Change the setup location Make the necessary adjustment via the scale operating menu see Menu Settings The weight readout...

Page 23: ...for a relatively long time under unfavorable conditions In this case notify the Sartorius Service Center Maintenance and repair work may only be performed by authorized service technicians who have ac...

Page 24: ...ntenance Do not apply any cleaning agents to ID labels or printed surfaces Corrosive Environment t Remove all traces of corrosive substances from the device on a regular basis Storage and Shipping Con...

Page 25: ...o http www sartorius com 2 Select the Services tab 3 Then select Disposal Information 4 Addresses for the local Sartorius disposal contacts can be found in the PDF files available for download on this...

Page 26: ...ve data Pin 3 TXD Transmit data Pin 4 DTR Data terminal ready Pin 5 GND Ground Pin 6 not used Pin 8 Clear to send CTS Make sure to observe the safety instructions t Remove the protective cover from th...

Page 27: ...evice ID explosion protection II 2G Ex ia IIC T4 Gb DEKRA 12ATEX0180 X The terminal control panel is suitable for use in potentially explosive areas in accordance with Directive 94 9 EC Zone1 devices...

Page 28: ...plosive area YPS02 XUR FM certified for the USA and Canada AC adapter for use outside of potentially explosive area YPS02 ZKR CSA certified for Canada AC adapter for use inside of potentially explosiv...

Page 29: ...Operating Instructions PMA 35001 X 29 EC Declaration of Conformity...

Page 30: ...30 Operating Instructions PMA 35001 X ECDeclaration of Conformity...

Page 31: ...it against any further use 10 All metal parts must be electrically connected to the terminal for the equipotential bonding conductor PA The equipment operator must to connect a lead with a gauge of at...

Page 32: ...32 Operating Instructions PMA 35001 X EC Verification...

Page 33: ...Operating Instructions PMA 35001 X 33 EC Verification...

Page 34: ...34 Operating Instructions PMA 35001 X EC Verification...

Page 35: ...02 X Circuit Uo Io Po Co Lo V_1 12 6 V 133 mA 1 43 W 0 5 F 300 H V_2 12 6 V 133 mA 1 43 W 0 5 F 300 H V_3 8 6 V 186 mA 1 49 W 0 8 F 200 H V_4 12 6 V 150 mA 1 65 W 0 5 F 300 H Alternative connections S...

Page 36: ...0 mH V_3 8 6 V 187 mA 1 51 W 391 nF 0 mH V_4 12 6 V 150 mA 1 68 W 223 nF 82 H 2 Power Supply YPSC01 X YPSC01 Z Circuit Uo Io Po Co Lo V_1 12 6 V 115 mA 1 45 W 0 5 F 350 H V_2 12 6 V 115 mA 1 45 W 0 5...

Page 37: ...b IIC or YCO01 Y interface converter II 2 GD EEx ib IIC or II 3 2 GD EEx nR ib IIC T4 FC FCA IS X 3 scale weighing platform see remark 5 on page 3 Zone 1 Gas IIC T4 Hazardous Area Non Hazardous Area Y...

Page 38: ...Co 13 5 F Ci 1 nF 8 4 F Lo 0 7 mH Li 1 6 H 0 7 mH Lo Ro 36 H 36 H Up to 7 additional CIXS3 or SIWS ISXS 3 Data cable Recommended Sartorius cable YCC485 X with approx 10 H ohm and 120pF m wire wire up...

Page 39: ...6 V 8 0 V 25 2 V Io 207 mA Ii 1 5 A Po 330 mW Pi 2 5 W Co 13 5 F Ci 1 nF 8 4 F Lo 0 7 mH Li 1 6 H 0 7 mH Lo Ro 36 H 36 H RS422 data interface 2 COM1 Uo 5 2 V Ui 8 6 V Io 290 mA Ii 210 mA Po 496 mW Pi...

Page 40: ...40 Operating Instructions PMA 35001 X Certificate of Compliance...

Page 41: ...Operating Instructions PMA 35001 X 41 Certificate of Compliance...

Page 42: ...42 Operating Instructions PMA 35001 X Certificate of Compliance...

Page 43: ...Operating Instructions PMA 35001 X 43 Certificate of Compliance...

Page 44: ...orms are used to facilitate legibility in these instructions and always simultaneously denote the other gender as well Copyright notice This instruction manual including all of its components is prote...

Reviews: