SARTORIUS PMA 7500...
11
Error Messages
Display
Cause
Solution
H
Scale capacity exceeded
Unload the scale
L
No pan on the scale
Place the pan on the scale
Something is touching the
Remove whatever is touching
pan
the pan
E 01
Display overflow; i.e., the value to
Reduce the load on the scale
be output cannot be displayed
E 02
Calibration parameter not met;
Only start calibration when zero
e.g.:
is displayed
– scale not zeroed
Press
=
to zero the scale
– load on scale
Unload the scale
E 08
Scale is not within the zero range
Scale can only be zeroed setting
within the zero-setting range
(± 2% of max. load)
E 09
Scale cannot be tared when the
Press
=
to zero the scale
gross value is
≤
0
E 10
=
key is blocked when there is
Press
c
to clear the 2nd tare
data stored in the 2nd tare memory
memory
E 11
Value cannot be stored in the
Press
=
2nd tare memory
E 22
Weight too light, or no load
Increase the weight on the scale
on the scale
E 30
p
key pressed while BPI mode
Set the scale to the SBI mode
active
(e.g., by activating the “Close”
function in the MC1 server)
E 50
Overflow or underflow in the
TK sensor or main PCB is defective
temperature compensation converter
(replacement required)
E 53
Temperature compensation converter
TK sensor or main PCB is defective
is not functioning. The processor
(replacement required)
is not receiving any measured data
from the TK switch
E 54
A/D converter output is below
The scale is underloaded; either
the minimum limit
the pan is not on the scale
or there is a mechanical defect
or a defect in the main PCB
E 55
A/D converter output is above the
Either the scale is overloaded
maximum limit
or there is a mechanical defect
or a defect in the main PCB