background image

4

General Information

Important:

Prerequisites for working on the Sartorius PMA Scales include considerable
experience, completion of a Service Course at the Training Center in
Goettingen, and the special tools and equipment described below.  Please
do not attempt or permit any unauthorized repair work!

If you remove the seal (warranty sticker), the scale is no longer covered
under the warranty.

Overview of Models

The Sartorius PMA7500... models are available in different versions:

PMA7500

Standard model

...D...

Dual Range Resolution (0g - 800g = 0,05g resolution,
>800g = 0,1g resolutiuon)

...X...

EX proof version

...C

with recalculation application (display head with 4 additional
keys)

Models

PMA7500, PMA7500D-000AM, PMA7500-000C, PMA7500-000W,
PMA7500-X, PMA7500-X00AM, PMA7500-X00C, PMA7500-X00W.
PMA7500D, PMA7500D-000AM, PMA7500D-000C, PMA7500D-000W,
PMA7500D-X, PMA7500D-X00AM, PMA7500D-X00C,
PMA7500D-X00W.
PMA7500D-X00CV1 (USA)

Differences to former PMA Versions

– new design
– new electronics
– higher capacity
– "Dual Range" resolution

Auxiliary Service Tools and Equipment

You will need the service software for adjusting overload stops, linearity and
span adjustment if the error in any of these adjustments is > 2%, and for
resetting the display and control unit after replacing one or more PCBs.

In addition to standard tools, you will need the following special tools:

Qty. Designation

Order no.

1

Sartorius service software, version 1.33 or later

6740-33

or

1

PSION Server, complete, version 4.6 or later

6739-98

1

RS 232 cable for connecting the scale to PC

7357 312

1

BPI-mode adapter

6740-63

Accompanying Literature

Installation and Operating Instructions for PMA7500... scales

Summary of Contents for PMA7500D-X

Page 1: ...WPM5003 e98061 Sartorius PMA7500 Electronic Paint Mixing Scales Service Manual for PMA7500 Models incl Spare Parts List and Service Specifications Sartorius AG Weighing Technology F...

Page 2: ......

Page 3: ...ng the Scale Operating Menu with or without Storing the New Settings 6 Activating the BPI Mode 6 Calibrating Adjusting the Scale 7 Calibration Adjustment Sequence 7 Zero Point Offset Value 7 Off Cente...

Page 4: ...PMA7500 X PMA7500 X00AM PMA7500 X00C PMA7500 X00W PMA7500D PMA7500D 000AM PMA7500D 000C PMA7500D 000W PMA7500D X PMA7500D X00AM PMA7500D X00C PMA7500D X00W PMA7500D X00CV1 USA Differences to former PM...

Page 5: ...code 8 1 2 If menu access is blocked this is indicated by a minus sign preceding the menu code number e g 1 1 1 1 1 rather than 1 1 1 1 1 To make the menu accessible select the menu code 8 1 1 the onl...

Page 6: ...ment error is 2 or replacing the main PCB Important After working in the BPI mode make sure to set the write protect again using the Close function so that the scale returns to the standard data trans...

Page 7: ...in 1 measuring voltage OP output IC6 pin 6 pin 2 ground Replace the pan support on the scale and fasten it with the screws place the pan on the pan support and re connect the scale to the power supply...

Page 8: ...lign the pan support arms approximately with the off center load bolts and place the pan on the pan support Check the off center load see Checking the Off Center Load above and adjust if necessary Lin...

Page 9: ...he old PCB change the solder bridges on the new PCB if necessary The menu access switch on the new PCB must be in the accessible position this is the position when the switch is moved toward the outer...

Page 10: ...head together with the connecting cable can be removed Follow the above instructions in reverse order to install the new PCB Replacing the Strain Gauge Load Cell Open the scale see page 9 Unplug the...

Page 11: ...data stored in the 2nd tare memory memory E 11 Value cannot be stored in the Press 2nd tare memory E 22 Weight too light or no load Increase the weight on the scale on the scale E 30 p key pressed wh...

Page 12: ...idual values 5 Interface 5 1 Baud rate 5 1 1 150 baud 5 1 2 300 baud 5 1 3 600 baud 5 1 4 1 200 baud 5 1 5 2 400 baud 5 1 6 4 800 baud 5 1 7 9 600 baud 5 2 Parity 5 2 1 Mark 5 2 2 Space 5 2 3 Odd 5 2...

Page 13: ...ance ppm K PMA 7500 7500 g 0 1 g 5000 g 0 1 g 5000 g 0 6 g 10 PMA7500D 999 95 7500 g 0 05 0 1 g 5000 g 0 1 g 5000 g 0 1 g 10 Span Linearity Model Class Adjustm weight Test weight Permissible Tolerance...

Page 14: ...South Afrika 1 1 1 6971411 Power supply STNG 6 240V Australien 1 1 1 6971412 Power supply STNG 6 Europa 230V 1 1 1 6971413 Power supply STNG 6 USA CDN 120V 1 1 1 6971414 Power supply STNG 6 240V GB 1...

Page 15: ...SARTORIUS PMA 7500 15...

Page 16: ...ns without the prior written permission of Sartorius AG The status of the information specifications and illustrations in this manual is indicated by the date given below Sartorius AG reserves the rig...

Reviews: