background image

  Assigning the 

I

 Toggle Key

  By pressing the toggle key 

I

 (6), the scale toggles between the basic setting (see page 15) and the  

settings defined under “SETUP- APPLICATION – UNIT” and “SETUP- APPLICATION – DECIMALS”

  Level 1    Level 2    Level 3    Level 4

 

SETUP  

 

 

 

 

 

 

 

 

 

 

 

 

 

§

  Press the 

L

 key [ENTER]  

 

 

 

APPLICATION   

 

 

 

 

 

 

§

  Press the 

RS

 keys to select  

 

               “APPLICATION” 

 

  

    

 

UNIT   

 

 

 

 

 

§

  Press the 

L

 key [ENTER] to select the 

S

  

 

               key 

“UNIT”; 

 

               press 

L

 key [ENTER]. 

 

      

 

 

 

 

PT./LB.   

§ 

Press the 

RS 

keys to select “GRAMS”,  

 

               for 

example 

 

 

 

 

 

 

 

 

 

 

 

o

  GRAMS

  

 

§

  Press the 

L

 key [ENTER], “o” appears;  

 

               the 

desired 

setting 

is 

defined 

 

              

§

  Press the 

c

 key (Clear) 

 

      

DECIMALS

   

 

  §

  Press the 

RS

 keys to select  

 

               “DECIMALS” 

 

 

          

  STANDARD 

§

  Press the 

L

 key; select setting

 

 

 

 

 

 

 

 

 

 

 

o

  POLYRANGE

 

§

  Press the 

L

 key [ENTER]; “o” appears,  

 

               Press 

the 

c

 key (Clear) to exit the menu

  Activating the “LOCK” Function “ ” 

  By activating the “LOCK” function, you can protect the scale from unauthorized use. When the scale 

is connected to a PC, the two devices are in constant communication. If the “LOCK” function is  

activated under “EXTRAS,” and data transmission to the PC is interrupted, the lock symbol will be 

displayed. The scale will automatically be locked, preventing other weighing operations. 

  Level 1    Level 2     Level 3     Level 4
 

SETUP

 

              

§

  Press the 

L

 key [ENTER]  

 

  

EXTRAS 

         

§

  Press the 

RS

 keys to select “EXTRAS”  

 

  

 

 

 

 

 

 

 

 

        §

  Press the 

L

 key [ENTER]  

 

      

LOCK 

      

§

  Press the 

RS

 keys to select “LOCK”  

 

        

  

 

  

    §

 Press the 

L

 key [ENTER]  

 

          

 

 

OFF 

§

  Select “ON” using the 

RS

 keys 

 

           o 

ON

     Confirm with the 

L

 key 

 

              

§

  Press the 

c

 key (Clear) several times 

 

               to 

exit 

the 

menu

  Entering a Password

  In addition to activating the “LOCK" function, the user may also enter a password.

  Should the user wish to deactivate the “LOCK" function by pressing the “OFF" key, he must first 

enter the valid password. The password is comprised of a 6-character numeric code.   

Use the 

RS

 keys to call up numbers (0 to 9). 

Six dashes ( ------ ) will appear in the display (10). The first dash will “blink" in the display. 

z

  

Select a number (0 to 9) using the 

RS

 keys; press the 

L

 key [ENTER] to save the number.   

The second dash will start to “blink." Repeat the aforementioned process. Should you wish to assign 

a “blank space," to one of the six characters, simply press the 

L

 key [ENTER] when the dash begins 

to blink. Once you have entered all six characters , simply press the 

L

 key [ENTER] when the cursor 

flashes in order to save the code.

16

Summary of Contents for PMA7501-Y00G

Page 1: ...98648 015 13 Installation Instructions Sartorius PMA World PMA Net Models PMA7501 Y00G Y00GL Electronic Paint mixing Scales for Use in Hazardous Areas Zone 2...

Page 2: ...or paint mixing applications 8 c key Clear and REC key for paint mixing applications 9 L key ENTER and MEM key for paint mixing applications 10 Graphic display 11 Weighing pan 12 Interfaces D Sub plug...

Page 3: ...t in a potentially explosive atmosphere Zone 2 make sure you observe the warnings and safety informa tion in their entirety during installation and operation as well as while perform ing maintenance a...

Page 4: ...2 Connect all equipment including peripheral devices to the equipotential bonding conductor Do not expose the scale to extreme tem peratures aggressive chemical vapors moisture shocks and vibrations T...

Page 5: ...ousing is made of PVC or rubber The allowable operating temperature range during operation is 0 C to 40 C 32 F to 104 F Make sure the place of installation is adequately ventilated to prevent build up...

Page 6: ...crew Strain relief for the Ethernet cable PMA7501 Y00GL only 2 protective plugs in the column Setting Up the Scale Choose a suitable place to set up the scale Avoid exposure to drafts heat moisture an...

Page 7: ...r supply the laws valid in your country must be observed If you should have any questions please contact your supplier or Sartorius Customer Service for information on the legal regu lations applicabl...

Page 8: ...til the desired weight of your formula is reached Remove the filled paint can from the weighing pan Never use a hammer to close the lid of a paint can while it is still on the weighing pan Otherwise y...

Page 9: ...mbol appears in this case g Pour in additional components until the desired weight of your formula is reached Remove the filled paint can from the weighing pan Important Note If you zero the display b...

Page 10: ...ant Note The flashing arrow n on the display means that the weight shown is not verified for use in legal metrology not legal for trade Example As you pour in the components of your formula the weight...

Page 11: ...r downwards using the S key 4 When you then press the L key MEM 9 the scale will automatically calculate and display the amounts of paint in g to add for each of the other components that you already...

Page 12: ...he display 9 Press the y key 4 several times to correct the value to 200 0 g 10 Press the L key MEM 9 COR 01 11 1 Add the first component C1 is displayed 1 7 g 12 Pour in paint until 0 0 g is displaye...

Page 13: ...onnection to AC power and before each calibration adjustment allow the scale to warm up for approx 30 min Hold down the U key 5 for 2 sec When 5000 is displayed release the key Center the calibration...

Page 14: ...d Use the RS keys to select the desired menu item here Ambient Conditions Press the L key ENTER to select the fourth level Level 4 Call up the menu item desired in the fourth level Use the RS keys to...

Page 15: ...ss the L key ENTER UNIT RS keys select DECIMALS GRAMS Press the L key ENTER o PT LB RS keys select STANDARD DECIMALS Press the L key ENTER o appears o STANDARD the desired setting is defined POLYRANGE...

Page 16: ...ol will be displayed The scale will automatically be locked preventing other weighing operations Level 1 Level 2 Level 3 Level 4 SETUP Press the L key ENTER EXTRAS Press the RS keys to select EXTRAS P...

Page 17: ...rd PW OLD will be displayed Following the correct input PW NEW will automatically appear You can now enter a new password or confirm each of the six blinking dashes by pressing the L key ENTER Blank s...

Page 18: ...ss the L key ENTER o will appear the factory settings are restored MENU appears on the display Press the c key Clear several times to exit the menu Setting Codes Under the setting CODES the menu items...

Page 19: ...other area changes constantly Too much vibration or the Access the menu to select scale is exposed to draft the appropriate code to adapt the scale to the particular weighing environment refer to Menu...

Page 20: ...e damage to the AC adapt er power supply or cables The equipment no longer functions properly The equipment has been stored for a relatively long period under unfavorable conditions The equipment has...

Page 21: ...information about disposal and recycling options please contact your local service staff The partners listed on the following website can be used for disposals within the EU 1 Go to http www sartorius...

Page 22: ...e Data Interface PMA Net RJ 45 10Base T 100Base TX Factory setting IP address 192 168 140 016 Gateway 192 168 140 001 Subnet mask 255 255 255 000 Port 21554 Specifications Model PMA7501 Y00G Y00GL Wei...

Page 23: ...ct the data cables YCO12 YCO13 when no power is being supplied to either end of the respective cables Securely fasten the D SUB connector to the PMA7501 Y The data cables YCO12 YCO13 may only be utili...

Page 24: ...24 Documents...

Page 25: ...25...

Page 26: ...26...

Page 27: ...27...

Page 28: ...28...

Page 29: ...29...

Page 30: ...30...

Page 31: ...31...

Page 32: ...32...

Page 33: ...33...

Page 34: ...ents GmbH Co KG Goettingen Germany All rights reserved No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius The stat...

Reviews: