22
Using the Balance
User Manual Secura
®
, Quintix
®
, Practum
®
5.2.2 Leveling with a Conventional Level Indicator
(Quintix
®
and Practum
®
)
L
R
R
L
L
R
The
Quintix
®
and Practum
®
models are fitted with a conventional level indicator .
This is located to the left of the display . The position of the air bubble in the indicator
shows whether or not the balance is correctly leveled .
t
Twist both of the front leveling feet of the balance to move the air bubble so that
it is centered in the circular marking .
The figure shows which leveling foot should be twisted in which direction . Generally,
both feet must be adjusted in order to position the air bubble exactly in the center of
the circular marking .
Examples:
– If the air bubble is too far to the lower right, twist the right leveling foot
counterclockwise (see figure) .
– If the air bubble is too far to the upper right, twist the left leveling foot
clockwise .
– If the air bubble is too far to the left, twist the left leveling foot counterclockwise
and the right leveling foot clockwise .
t
Check the level indicator after each adjustment and correct if necessary .
t
Complete the leveling by calibrating the balance .