7.3 Individual Identifiers (Secura
®
only)
Purpose: You can define identifiers for the following print jobs:
– Device identification (ID): printed in header of GLP printout .
– Additional ID (A ID): printed in header of GLP printout .
– Batch ID (L ID): queried after each GLP head in GLP printout at beginning of
measurement .
– Sample ID (S ID): for printouts via the
g
button . This ID is queried before each
printout .
Prerequisites: – The balance must be connected to a Sartorius lab printer or PC for printouts (see
“USB Port” on page 79) .
– The settings for individual identifiers can be configured in the system settings
under
Identifier
(see “Identifier (Secura
®
only)” on page 35) .
– The GLP printout can be activated in the system settings under
Print function
in
Manual print
(see “Printout” on page 34) .
Starting the Printout
Example: Configuration of system settings for printing sample ID (S ID):
Identifier
–
Sample ID function
–
Autom. increment
:
t
Place the sample on the weighing pan .
14.42
Max 2100 g
d=0.01 g
+
t
Select
g
to print .
1
S ID
1
2
3
4
5
6
7
8
9
-
0
.
C
y
The identifier for the printout is displayed .
The displayed value can be changed if desired .
– Delete the last character: Select
k
.
t
Select
l
to confirm .
y
The weight value is printed . A sample printout can be seen below:
------------------
S ID
11
N
+
200.21 g
S ID
12
N
+
200.19 g
Summary of Contents for Practum
Page 1: ...98648 019 43 User Manual Secura Quintix Practum Laboratory Balances 98648 019 43...
Page 2: ......
Page 117: ...117 Specifications User Manual Secura Quintix Practum EC Type approval Certificate...
Page 118: ...118 Specifications User Manual Secura Quintix Practum Plates and Markings...
Page 119: ...119 Specifications User Manual Secura Quintix Practum Plates and Markings...