49
Weighing
User Manual Secura
®
, Quintix
®
, Practum
®
7.5 Statistics (Secura
®
and Quintix
®
only)
Purpose: Save weight values and statistically evaluate them . You can save up to 99
components . The values are generated as results:
– Number of components
– Average
– Standard deviation
– Variation coefficient
– Sum of all values
– Lowest value (minimum)
– Highest value (maximum)
– Spread: Difference between maximum and minimum
Prerequisites: For printouts, a printer must be connected and configured (see “USB Port” on
page 79) .
t
Select the Menu key in any application .
t
Select the
Statistics
symbol in the menu .
Max 2100 g
d=0.01 g
0.00
Component
1
y
The
Statistics
application appears .
Component
1
t
To change settings, select the gray button .
Statistics
Print component
Automatic tare
Off
On
On
Off
y
The
Statistics
settings window appears .
t
Choose whether the individual weights of the components are printed out .
t
Determine whether the balance is tared when a component is saved .
t
Select
l
to confirm .
t
Select
A
to zero the balance if necessary .
t
Place a container on the weighing pan .
Summary of Contents for Practum
Page 1: ...98648 019 43 User Manual Secura Quintix Practum Laboratory Balances 98648 019 43...
Page 2: ......
Page 117: ...117 Specifications User Manual Secura Quintix Practum EC Type approval Certificate...
Page 118: ...118 Specifications User Manual Secura Quintix Practum Plates and Markings...
Page 119: ...119 Specifications User Manual Secura Quintix Practum Plates and Markings...