Activating a Data Output Process
You can define the data output parameter so that output is activated either
automatically or when a print command is received. You have two options for the
automatic mode: data output can be either synchronous with the scale display
or activated at defined intervals (to select the parameter, see “Utilities for Printouts
or Data Transfer”).
Data Output by Print Command
The print command can be transmitted by a software command or by pressing the
corresponding key.
In addition to an interface cable for a different device, you can connect an
external universal switch for remote control to the scale interface port using the
YCC01-0016M3 adapter cable (for the print function, see Part 2). For the switch,
use pins 8 (ground) and 15 of this port and a cable up to 1.5 m or 5 ft. long
(RS-232C).
Use pins e (ground) and m on the 12-pin round connector.
If data output is requested by a software command (see the section on “Data Input
Formats”), you can install a 15 m (50 ft) cable for RS-232C, or a 300 m (984 ft)
cable for RS-423.
Automatic Data Output
In the “auto print” operating mode, the data are output to the interface port
without requiring a print command. You can choose to have data output
automatically at defined print intervals with or without the stability parameter.
If you select the auto print setting, data will be transmitted immediately the
moment you turn on the scale. This data output function is described in Part 2,
under “Utilities for Printouts or Data Transfer.”
4–15