Changing the Reference Sample Quantity “
nRef
”
You can change the reference sample quantity in cycles by pressing
g
/
W
key.
Use the number keys
0
–
9
and the
u
/
w
key to enter piece counts.
Choose from the following settings for the reference sample quantity: 5, 10, 20,
50, 100 and a user-definable number (factory setting: 999).
To change the ref. sample aty: press the
g
/
W
key
(13)
repeatedly and
select the desired reference sample quantity, or
enter directly using the numeric keys
0
–
9
.
After the reference sample quantity “100 pieces,” 999 (or another number) is
displayed. You can overwrite this number directly using the numeric keys
0
–
9
(saved in non-volatile memory).
Note: This setting is
not
canceled by the reset code 9 – – 1°!
If the scale is not loaded with the corresponding number of pieces, the “Err 22”
error code is displayed.
Make sure you enter the number for “reference sample quantity ‘
nRef
’” without
the decimal point (
.
key)!
A number entered manually without a decimal point is always a piece count!
You can exit this application at any time by pressing
c
/
c
+
u
/
w
or
g
/
W
, and proceed with a new counting task.
Input of Reference Weight “wRef”
If you know the weight of a part, you can enter it directly as the reference weight
“
wRef
” using the numeric keys
0
–
9
.
Make sure you enter the number with the decimal point (
.
key)!
A number entered manually with a decimal point is always a weight value!
You can enter a number of up to 7 digits manually.
Display reference weight for 2 seconds when
the counting function is started by pressing
u
/
w
Code
Changes
Blocked
3
9
1 *
Display
3
9
2
* = factory setting