25
Service Manual Signum
Working with the Service Mode
– Press the
J
key to store the 3
rd
linearization weight. The 4
th
required weight is
now displayed.
– Place the required weight on the scale. After a brief pause, the difference
between the current value and the value from the last adjustment (calibration)
procedure is displayed
– Press the
J
key to store the 4
th
linearization weight. The scale now prompts a
zero point (remove all weights from the scale).
– Once the zero point is stored, the scale will automatically return to the weighing
mode.
*)
Weights displayed depend on the scale model.
3)
Only displayed if menu code »
1-10-2
« (
Calibration with manual adjustment) has been selected.
10kg.eps
display_serv_06.eps
display_serv_04.eps
0kg.eps
*
*
3)