27
Service Manual Signum
Working with the Service Mode
– After approx. 2 seconds, the 1
st
required linearization weight (user-defined
weight) is displayed. You can enter the linearization weights via menus »
1-18-2
«
through »
1-18-5
«
2
.
– Place the required weight on the scale. After a brief pause, the difference
between the current value and the value from the last adjustment (calibration)
procedure is displayed
– Press the
(
key if you need to interrupt the calibration/adjustment procedure.
– Press the
J
key to store the 1
st
linearization weight. Next, the 2
nd
required
weight is displayed.
– After a brief pause, the difference between the current value and the value from
the last adjustment (calibration) procedure is displayed
– Press the
J
key to store the 2
nd
linearization weight. Now the 3
rd
required
weight is displayed.
– After a brief pause, the difference between the current value and the value from
the last adjustment (calibration) procedure is displayed.
*)
Weights displayed depend on the scale model
3)
Only displayed if menu code »
1-10-2
« (
Calibration with manual adjustment
) has been selected.
– Press the
J
key to store the 3
rd
linearization weight. The 4
th
required weight is
2.5kg.eps
display_serv_02.eps
5kg.eps
display_serv_03.eps
7.5kg.eps
display_serv_04.eps
*
*
*
3)
3)
3)