33
Service Manual Signum
Working with the Service Mode
1-18-2 Entering the 1
st
Linearization Weight
– Scroll to menu code »
1-18-2
« (by pressing the
k
key).
Important note:
The service mode must be active.
– Access the input mode for the 1
st
linearization weight by briefly pressing the
)
key. The first number blinks.
– Enter the 1
st
linearization weight (2,500 kg in this example); press the following
keys in the order indicated: 3*
k ) p )
6*
k )
k ) k
. Now
the last number entered blinks.
– Press the
J
key to store the 1
st
linearization weight. Then the scale returns to
menu code setting »
1-18-2
«.
– Using the
k
key, scroll to the menu code settings »
1-18-3
«, »
1-18-4
«, and
»
1-18-5
« and enter the values for Lin. 2, Lin. 3 and Lin. 4 according to the
instructions above.
If you do not require all linearity positions, enter „0“ in the unused fields to blank
these lines in the display.
– Once all entries have been made, menu code setting »
1-18-5
« must be
displayed.
menu_1182.eps
user1.eps
userl2.eps
menu_1182.eps
menu_1185.eps