Service Manual Signum
Error Codes
Display
Cause
Solution
Inf 07
Function not allowed in scales
Contact your local Sartorius Service Center
verified for use in legal metrology
for details on changing settings
Inf 08
The load on the scale is too heavy to zero the readout Check whether “Tare/zero at power on” is set (1.12)
Inf 09
Taring is not possible when the gross weight
Zero the scale
is a minus value
Inf 10
Tare key is blocked when there
The data stored in the 2nd tare memory
is data in the tare memory
(Combics 2 only) must be deleted (clear
the memory) before taring
Inf 22
Error in storing reference value, load is too light
Put a heavier weight on the scale
Inf 23
Error in initializing an application
Contact your local Sartorius Service Center
Inf 29
Minimum load not reached
Change menu, or perform “Close” function check
the interface and cable of the connected devicen
Contact your local Sartorius Service Center
Inf 30, 31
Indicator is in the xBPI-mode
Define a lower value for the minimum load
(in the Application settings, menu item 3.6)
Inf 71
Cannot store the current weight value
None
(e.g., if control limits are too low or too high)
Inf 72
Cannot store the current weight value
None
(e.g., the transaction counter has reached its limit)
Inf 73
Data not found or unreadable
Contact your local Sartorius Service Center