37
Adjusting Linearity
Checking the Linearity
Note:
The linearity must be checked according to the Standard Operating Procedure
WKD-038.
- Check the linearity of the balance/scale for the entire weighing range in 4-g
steps.
- Compare the values displayed with the tolerance ranges given in the „Table
of Calibration/Adjustment Data“ on pages 53-54.
- If the erors exceed the permissible tolerance ranges, the linearity must be
adjusted
Adjusting the Linearity
Use the service software to adjust linearity.
– Activate the BPI mode (see page 26).
– Connect the balance/scale to the PC and call the linearity program in the
service software.
– Test the span and adjust if necessary (see page 36, „Span Adjustment“).
– After linearity and span has been adjusted, activate the „Close“ function in
the service software to reset the balance/scale to the SBI mode.