41
Checking the Off-Center Error
Note:
The data for the test weight and the tolerance are listed in the „Adjustment
Datasheet (see Page 53-54).
CP Balances/Scales with Draft Shields and Triangular Weighing Pans
Models: TE64; TE124S; TE214S;TE313S; TE313S-DS; GD103; GD603;
- Place the test weight on position 1 on the weighing pan and zero the
balance/scale by pressing the
t
key.
- In the order given, place the test weight in the position 2, 3, 4 and 5 as
indicated and write down the readout at stability, including the plus/minus
signs.
- If the off-center loading error is too large, adjust the balance so that the
off-center loading error is within the tolerances specified.
TE- Balances/Scales with Square Weighing Pan
Models: TE3102S
- Compare the off-center loading errors with the tolerances listed in the
(„Adjustment Datasheet“ see Page 53-54).
- If the off-center loading error is too large, adjust the balance so that the
off-center loading error is within the tolerances specifie
eck_ru_03.eps
eck_re_03.eps
5
2
1
3
4
5
2
1
3
4
Summary of Contents for Talent Series
Page 55: ...55 ...