Procedure for Adjusting the Off-Center Error (Double Lever
System)
Note:
After calibration/adjustment of the off-center loading error, you must check
the span and linearity and adjust as required.
On models TE64;TE124S;TE214S;TE313S;TE313S-DS;
GD103;GD603; (see upper illustration on the left) TE3102S (see lower
illustration on the left) the off-center loading error should be adjusted by
careful filing at appropriately thin places on the upper guide. To avoid
distorting the adjustment, wait a few seconds after filing so that the thin
sites can „cool down“.
Caution!
After carefully removing the housing, adjust the off-center
loading error only at the thin sites in the back using an angled
file.
Check the off-center loading error and adjust the largest error
first by carefully filing at these points
(see illustrations on page 45).
The errors measured (see the section „Off-Center Loading Error) refer to the
position of the systems in the balance/scale.
Aut12954b.jpg