32
Interface
D
escription (Option)
lf you wish to record weight data using the Sartorius Data Printer, just plug the printer
connector Mo the interface port of the scale. You do not need to adjust any settings!
(You can also interface your scale with a Computer.)
General Specifications
Type of interface
Serial point-to-point connector
Operating mode
Asynchronous, simplex
Standard
V24-V28, RS232 C-S
Handshake lines
Clear to Send (CTS)
Data Terminal Ready (DTR)
Initialization
External or automatic print command depending
of the interface
on the code selected (8 1 to 8 16)
Character coding
7-bit ASCII
Transmission rate
1,200 baud
Parity Odd
Synchronization
1 Start bit, 1 stop bit
Additional port lines for
4, for upper/lower limits,
over/under checkweighing
tolerance range and target
Data Output Format
1 2 3 4 5 6 7 8 9 10 11
12 13 14 15
16 17 18 19 20 21 22
H H - - - -
+
-
- - Z Z Z Z Z Z Z _ X X X CR
LF
1st and 2nd characters:
Weight difference or net weight
3rd to 6th characters:
Space
7th character:
Plus or minus signor space
8th and 9th characters:
Space
10th to 16th characters:
Digit, space or decimal point
17th character:
Space
18th to 20th characters:
Letter or space
21st character:
Carriage Return
22nd character:
Line Feed