36
Specifications
Model
TS 6100B
TS 12
Capacity g
6,100
12,100
Readability g
0.5
1
Tare range (by subtraction)
g
-6,100
-12,100
Standard deviation
g
≤
± 0.5
≤
± 1
Max. linearity
g
≤
± 0.5
≤
± 1
Stabilization time (typical)
s
1.1
Display Update
– when the readout stabilizes
(depends on the filter level
selected)
– when the load changes
s
s
0.1/0.2
0.1
Adaptation to application requirements
and ambient conditions
By selection of one of 2 optimized filter levels
Stability range
d
0.25 ... 32 (selectable)
Ambient temperature range
K
273 — 313 (0
o
C...+40
o
C) (32
o
F … 104
o
F)
Moisture-proof rating acc. to DIN 40040
Class F, non-condensing
Dust and water protection rating in
conformance with DIN 40050/IEC529
IP65
Sensitivity drift within 10...30°C
/
o
C
≤
±2.5
∙
10
-6
Platform size
mm
in.
260 x 210
10.2 x 8.4
Scale housing (W x D x H)
mm
in.
260 x 295 x 110/450
10.2 x 11.6 x 4.3/17.7
Net weight, approx.
kg
~lb
6.5
14.3
Power requirements
(v frequency)
115 or 230 depending on the type
of power supply used; 50-60 Hz
Allowable voltage fluctuation
-20% ... +15%
Power consumption (typ.)
VA
14
Summary of Contents for TS 12