background image

 

11 

 
 
 
 
After the balance operating program has been called the display 
will show a continuous numerical sequence from 0 to 5 repre-
senting the "page" selection, in addition to the Status code letter 
''L" or "C.'' When the code number of your choice for the "page" 
appears, press the tare control. The "page" code number now 
stops in the display, and a series of numbers for the "lines'' will 
begin to cycle. Again, press the tare control to stop the code 
number of your choice in the display. Next the numbers for the 
"word'' will cycle in the display. Repeat the procedure to enter 
the last digit of the code. 
 
The "o" symbol that appears indicates the actual setting. 
 
To change any settings ("C" mode), press the tare control as 
soon as the appropriate code appears. 
 
Brief display of BUSY and the "o" symbol confirms your selec-
tion, followed by a return to ''zero" representing the "line." 
 
 
To return to the weighing program: 
press the tare control each time a 0 appears in the numerical 
sequence (word, line, page). lf you have made code changes 
your code entry will be stored as soon as the display returns to 
the weighing mode. Lock the balance operating program using 
the menu access switch (display ''L") and replace the protective 
cap. 
 
 

Auto Zero 

 
This balance has an automatic zero tracking function, known as 
"Auto Zero" (can be turned off by menu code). 
Any changes off zero 

 0.5 of a digit per second will be set to 

zero automatically. 
 
 

 

Summary of Contents for U 4600P

Page 1: ...Sartorius Universal U 4600 P Electronic Precision toploader Installation and Operating Instructions...

Page 2: ...receptacle for line cord power lead 3 Leveling foot 4 Menu access switch 5 ON OFF key 6 CAL key 7 PRINT control functions only if balance has a built in interface 8 Tare control 9 Weight display 10 L...

Page 3: ...se installation and operating instructions care fully before operating your new toploader This equipment complies with the requirements in Part 15 of FCC Rules for a Class A computing device Operation...

Page 4: ...balance operating program For this purpose please refer to pages xx After you plug your toploader into a wall outlet allow for more than 30 minutes warmup Important Note Unplug your toploader before...

Page 5: ...miss out on the benefits of our warranty Please fill out the warranty card indicating the date of installa tion and return the card to your Sartorius dealer The equipment supplied comprises the compo...

Page 6: ...0 V make sure to exchange the installed T 80 mA fuse 2 for the T 160 mA fuse that comes with the toploader This fuse along with the voltage selector is plugged into the rear panel of the toploader 100...

Page 7: ...cessing a function and will not ac cept any other commands to perform functions at this time STANDBY The toploader has been switched off with the ON OFF key 5 and is now in the STANDBY mode POWER OFF...

Page 8: ...zed STANDBY This means the toploader is immediately ready to operate without requiring warmup the next time you switch it on After the power is turned on a test of all essential electronic functions i...

Page 9: ...Signal in dicates the end of the calibration procedure You can block the calibration function see Balance Operating Program The function is accessible whenever the balance op erating program is unloc...

Page 10: ...balance turned off STANDBY state hold down the tare control 8 and simultaneously press the ON OFF key 5 Upon completion of the automatic self test release the tare control as soon as CH 5 is displayed...

Page 11: ...mbol that appears indicates the actual setting To change any settings C mode press the tare control as soon as the appropriate code appears Brief display of BUSY and the o symbol confirms your selec t...

Page 12: ...2 Singapore taels tl C 1 3 1 Last decimal ON c 5 2 3 Taiwan taels tl C 1 3 2 Last decimal OFF c 5 2 4 Grains gr C 1 3 3 Last decimal at stability c 5 2 5 Pennyweights dwt C 1 3 4 All decimals at stab...

Page 13: ...s and application requirements by selection of one of four digital filter levels Stability range d 0 25 64 selectable Ambient temperature range K 273 313 0 o C 40 o C Sensitivity drift within 283 303...

Page 14: ...Copyright by Sartorius AG G ttingen Deutschland All rights reserved No part of this publication may be printed or translated in any form or by any means without the prior written permission of Sartori...

Reviews: