Operation. 

 
 
 
The weight display shows the following special Status mes-
sages for your information: 
 

BUSY 

The processor is still busy processing a function and will not ac-
cept any other commands to perform functions at this time. 
 

STANDBY 

The toploader has been switched off with the ON/OFF key 

(5)

 

and is now in the STANDBY mode. 
 

POWER OFF 

The toploader was disconnected from line power (reconnection 
to line power, power failure). 
 

CAL 

The calibration function has been called. 
 
 
 

Summary of Contents for U 4600P

Page 1: ...Sartorius Universal U 4600 P Electronic Precision toploader Installation and Operating Instructions...

Page 2: ...receptacle for line cord power lead 3 Leveling foot 4 Menu access switch 5 ON OFF key 6 CAL key 7 PRINT control functions only if balance has a built in interface 8 Tare control 9 Weight display 10 L...

Page 3: ...se installation and operating instructions care fully before operating your new toploader This equipment complies with the requirements in Part 15 of FCC Rules for a Class A computing device Operation...

Page 4: ...balance operating program For this purpose please refer to pages xx After you plug your toploader into a wall outlet allow for more than 30 minutes warmup Important Note Unplug your toploader before...

Page 5: ...miss out on the benefits of our warranty Please fill out the warranty card indicating the date of installa tion and return the card to your Sartorius dealer The equipment supplied comprises the compo...

Page 6: ...0 V make sure to exchange the installed T 80 mA fuse 2 for the T 160 mA fuse that comes with the toploader This fuse along with the voltage selector is plugged into the rear panel of the toploader 100...

Page 7: ...cessing a function and will not ac cept any other commands to perform functions at this time STANDBY The toploader has been switched off with the ON OFF key 5 and is now in the STANDBY mode POWER OFF...

Page 8: ...zed STANDBY This means the toploader is immediately ready to operate without requiring warmup the next time you switch it on After the power is turned on a test of all essential electronic functions i...

Page 9: ...Signal in dicates the end of the calibration procedure You can block the calibration function see Balance Operating Program The function is accessible whenever the balance op erating program is unloc...

Page 10: ...balance turned off STANDBY state hold down the tare control 8 and simultaneously press the ON OFF key 5 Upon completion of the automatic self test release the tare control as soon as CH 5 is displayed...

Page 11: ...mbol that appears indicates the actual setting To change any settings C mode press the tare control as soon as the appropriate code appears Brief display of BUSY and the o symbol confirms your selec t...

Page 12: ...2 Singapore taels tl C 1 3 1 Last decimal ON c 5 2 3 Taiwan taels tl C 1 3 2 Last decimal OFF c 5 2 4 Grains gr C 1 3 3 Last decimal at stability c 5 2 5 Pennyweights dwt C 1 3 4 All decimals at stab...

Page 13: ...s and application requirements by selection of one of four digital filter levels Stability range d 0 25 64 selectable Ambient temperature range K 273 313 0 o C 40 o C Sensitivity drift within 283 303...

Page 14: ...Copyright by Sartorius AG G ttingen Deutschland All rights reserved No part of this publication may be printed or translated in any form or by any means without the prior written permission of Sartori...

Reviews: