Technical Data
15 Technical Data
15.1 General Data
Specification
Unit
Value
Scale
Supply voltage
Only via Sartorius power supply YPS07-USB
Input voltage
V
DC
+5.0
Power consumption
W
5.1
Further data
IP40 in accordance with EN 60529/IEC 60529
Ambient conditions
The specifications apply under the following ambient conditions:
Environment
For indoor use only.
Operational capability
°C
Guaranteed b5 and +40.
Storage and shipping
°C
-10 to +60
Relative humidity
%
Up to 80% for temperatures up to 30°C non-condensing, decreasing
linearly up to 50% relative humidity at 40°C.
Ex-link converter interface connection
Ethernet
Electromagnetic Compatibility
In accordance with EN 61326-1/IEC61326-1 Electrical equipment for
measurement, control and laboratory use – EMC requirements – Part 1:
General Requirements.
Interference resistance
Basic requirements
Transient emissions
Class B
Suitable for use in residential areas and areas that are connected to a low
voltage network that also supplies residential buildings.
Verified scales in accordance with EU requirements comply with the
requirements of Council Directive 2014/31/EU with EN 45501:2015 and
OIML R76:2006.
* For verified scales in accordance with EU requirements, refer to the
information on the scale.
** For verified scales in accordance with EU requirements, the legal
regulations apply.
Available application programs
Recalculation, factor calculation, formula
Power supply YPS07-USB
USB power supply unit (5 V/1500 mA)
Typee FW7721M (manufacturer description)
Primary
100–240 V~, –10%/+10%, 50–60 Hz, 200 mA–100 mA
Secondary
5 V
DC
, ± 5%, 1500 mA (max.)
Further data
Protection class II
IP40 in accordance with EN 60529/IEC 60529
Ex-Link Converter YCO16-Z
Further data
IP40 in accordance with EN 60529/IEC 60529