66
VIS1X
Operating Instructions
Conformity & Approvals
To verify the availability of the Approved product, please refer to www.approvalguide.com
FM Approvals HLC 5/13
3055566
Page 1 of 3
FM Approvals
1151 Boston Providence Turnpike
P.O. Box 9102 Norwood, MA 02062 USA
T:
781 762 4300
F: 781-762-9375 www.fmapprovals.com
CERTIFICATE OF COMPLIANCE
HAZARDOUS (CLASSIFIED) LOCATION
ELECTRICAL EQUIPMENT
This certificate is issued for the following equipment:
PMA.Vision VIS1Xab-c
IS / I / 1 / CD / T4 - 2003809
I / 1 / Ex ia / IIB / T4 - 2003809
a = Accessories: Up to three letters and/or numbers or blank (not critical to safety)
b = Metrology Approval: Up to two letters and/or numbers or blank (not critical to safety)
c = Up to five letters and/or numbers or blank (not critical to safety)
Specific conditions of use
1. The PMA.Vision shall only be used with the YCO16-Z Ex-Link Converter.
2. The protective ground connectors on the PMA.Vision and the YCO16-Z Ex-Link Converter shall be connected together
using a potential equalization conductor with a cross-sectional area of at least 4 mm
2
.
3. The CAT6 cable used between the PMA.Vision and the YCO16-Z Ex-Link Converter shall be obtained from Sartorius.
4. The enclosure of the PMA.Vision is non-conducting and may generate an ignition-capable level of electrostatic charges
under certain extreme conditions. The user shall ensure that the equipment is not installed in a location where it may
be subjected to external conditions that might cause a build-up of electrostatic charges on non-conducting surfaces,
additionally; cleaning of the equipment should be done only with a damp cloth.
PMA.Vision VIS1Yab-c
NI / I / 2 / CD / T4 – NIFW; 2003809
a = Accessories: Up to three letters and/or numbers or blank (not critical to safety)
b = Metrology Approval: Up to two letters and/or numbers or blank (not critical to safety)
c = Up to five letters and/or numbers or blank (not critical to safety)
Specific conditions of use
1. The PMA.Vision shall only be used with the YCO16-Z Ex-Link Converter.
2. The protective ground connectors on the PMA.Vision and the YCO16-Z Ex-Link Converter shall be connected together
using a potential equalization conductor with a cross-sectional area of at least 4 mm
2
.
3. The CAT6 cable used between the PMA.Vision and the YCO16-Z Ex-Link Converter shall be obtained from Sartorius.
4. The enclosure of the PMA.Vision is non-conducting and may generate an ignition-capable level of electrostatic charges
under certain extreme conditions. The user shall ensure that the equipment is not installed in a location where it may
be subjected to external conditions that might cause a build-up of electrostatic charges on non-conducting surfaces,
additionally; cleaning of the equipment should be done only with a damp cloth.
YCO16-Z Ex Link Converter
AIS / I / 1 / CD – 2003809