X3 Operating Manual
Fieldbus Interface
Sartorius
133
7.3.6 Register 5: Device Type and Software Release
Only reading is permitted.
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
TYPE MSB
Byte 1
TYPE LSB
Byte 2
MAINVERSION
Byte 3
SUBVERSION
E.g. 5410 Rel 1.23 = 54100123
hex
7.3.7 Register 6: Board Number
Only reading is permitted.
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Board number MSB
Byte 1
""
Byte 2
""
Byte 3
Board number LSB
E.g. 148388723 = 08D83B73
hex
7.3.8 Register 7: (Reserved)
7.3.9 Register 8 ...15: Weight Data
Only reading is permitted.
The gross, net and tare weight are stored as DINT fixpoint.
The real data value is derived from DINT and EXPO as follows :
Value
Real
=reading
DINT
*10
(-EXPO
Register 8
Current gross value
Register 9
Current net value, if tared; otherwise gross
Register10
Current tare value, if tared, otherwise 0
Register11
Value on the front-panel display
Register12 Reserved
Register13 Reserved
Register14 Max
Register15 Reserved
7.3.10 Register 20: Weight Data
Register 20 BCD output value
7.3.11 Register 22 ... 27: Limit Values (Read/Write)
Register 22
Limit 1 on
Register 23
Limit 1 off
Register 24
Limit 2 on
Register 25
Limit 2 off
Register 26
Limit 3 on
Register 27
Limit 3 off
Summary of Contents for X3 PR 5410
Page 14: ...Process Indicator X3 Operating Manual 14 Sartorius 2 3 6 Overview of Accessories ...
Page 144: ......
Page 147: ......