X3 Operating Manual
Commissioning
Sartorius
75
4.5.8 Calibration with Load Cell Data (“Smart Calibration“)
This is the simplest method for 'calibration without weights', as no calculation is necessary.
WP A/Calibration
Max
3000 d
3000 kg
Scale interval
3000 d
1 kg
Deadload at
3.00 kg
0.001000
Max at
3000.00 kg
0.000000
Not calibrated
Sensitivity 833.33
4.000000
by load
by mV/V
by data
Linear.
Test
Start by pressing [by data].
WP A/Calibration/Loadcell configuration
Number of loadcells
4
Nominal load
3000 kg
Gravity
9.81379 m/s²
Hysteresis error
not specified
Certified data
all LC same
LC sensitivity
1.000000
Enter
Calc
WP A/Calibration/Loadcell configuration
LC resistance
600.000 ¥
Enter
Calc
[Number of loadcells]
Number of load cells connected in parallel (1, 2...<4>...9, 10)
[Nominal load]
Nominal load E
max
of a load cell (not the overall nominal weight of the scale).
[Gravity]
Acceleration of gravity at the place of installation; default is the value for Hamburg, Germany, 9.81379
m/s².
[Hysteresis error]
When switching from [Not specified] to [Specified], values for [Correction A/B] must be filled in. The data is
given on the load cell certificate.
[Certified data], [LC sensitivity], [LC resistance]
With [all LC same] only 1 value for the sensitivity [LC sensitivity] and the output resistance [LC resistance]
must be filled in. With [each LC specific] individual values for each load cell are requested.
[Calc]
The mV/V value is calculated and after confirmation with [OK], the calculated mV/V value is stored in the
calibration data.