X3 Operating Manual
Commissioning
Sartorius
77
4.5.11 Test Value Determination / Display
Press [Test] to activate test value calculation. The maximum capacity (Max) is displayed with the designation
without a weight unit.
WP A/Calibration
Test
Depending on the settings under -[Calibration]-[Param]-[Test mode], either
- the maximum capacity with [Absolute], or
- the deviation from the test value with [Relative]
is shown when you press
to view the test data.
4.5.12 Finishing / Saving the Calibration
Finish the calibration with .
You are prompted to confirm whether calibration should be closed without
determining the test value.
Calibration not complete
Exit calibration?
If not all data was determined when calibrating with [New] (e.g. dead load not
set/entered), this message is shown:
Press [Yes] to confirm and then press again; another prompt is displayed:
Exit calibration ?
Press [Save] to save changes in calibration data. If you press [Undo], changes
are not saved and the display returns to the selection menu for the weighing
points.
Saving calibration
Confirm, or
Exit calibration
close the menu.
After finishing the calibration, set the CAL switch to the closed position; see also Chapter 4.1.1.