X3 Operating Manual
Commissioning
Sartorius
85
4.7.5 Operating
Parameters
Define the basic operating parameters under -[
Operating parameter
].
Setup/Operating parameter
Address
A
PIN
******
Use alibimemory
none
Sequencenumber 0
SetTareKey
tare
&
reset tare
SetZeroKey
only when not tared
ShowGrossTarekey enabled
PrintKey enabled
Testkey enabled
[Address]
Device address; e.g. for print-out.
[PIN]
The access code can be used to protect the [Setup] from unauthorized operation. Enter a number with up to
6 digits. As long as you are in this menu, the value can be overwritten as required.
[Use alibimemory]
The dataset to be stored in the alibi memory can be defined:
[none] - nothing, [Gross] - Gross weight, [Net] - Net weight, [Gross,Net,Tare],
[Gross,Net], [Gross,Tare].
[Sequencenumber]
Number may appear (selectable) on the print-out too, is incremented automatically ( max. 999999 ) and can
be set here.
[SetTareKey]
The function of the tare key on the keypad can be configured:
[tare
&
reset tare]: the scale is tared if it was not tared previously; otherwise, tare is reset.
[tare
&
tare again]: when you press the [Tare] key, the current value is stored in the tare memory and a net
weight of 0 is displayed. [disabled]: The key has no function.
[SetZeroKey]
[only when not tared]: The function of the zero-setting key on the keypad can be limited to gross mode.
[reset tare on zeroset]: The zero-setting key switches the scale to gross mode automatically. If the zero-
setting key with these settings has no effect, the configured zero-setting range (around the zero-point set
with the dead load) is already utilized due to a previous zero-setting operation and/or automatic zero
setting. [disabled]: The key has no function.
[ShowGrossTarekey], [PrintKey], [Testkey]
The front panel keys for selecting between gross, net and tare, for starting a print command and for starting
the analog test can be de-activated by [disabled]. With [enabled], operation is possible.
LOCKEDw
Answer on the instrument display, if disabled keys are pressed.
Save changes ?
To close the menu, press . Press [Yes] to save the data, or [No] to close the
menu without changing data.