Commissioning
X3 Operating Manual
94
Sartorius
If the limits (Limit 1 and Limit
2) for 'On' and 'Off' are equal
(on = off), output 1 (Limit 1
out) switches ON, when the
weight (Wgt) exceeds the value
and output 2 (Limit 2 out)
switches OFF, when the weight
drops below the value.
Setup/Limit parameter
Limit 1 on
900 kg
Action
set marker 1
X64=1
Condition
no
condition
-----
1. Determining an action
Determine the action for the rising edge of the reference signal under [Limit 1 on] from the following list
(here: Marker 1 is set when 900 kg are exceeded).
Accordingly, an action for [Limit 1 off] can be determined.
Function
SPM bit
-no action-
-----
No function
set marker 1
X64=1
Set marker 1
set marker 2
X65=1
Set marker 2
set marker 3
X66=1
Set marker 3
select net
X72=1
Select net
set zero
X112=1
Set zero
set tare
X113=1
Set tare
reset tare
X114=1
Reset tare
set test
X115=1
Activate the analog test
reset test
X116=1
Finish the analog test
reset PWF
X117=1
Reset power fail
set fixtare
X118=1
Set fixtare (use the value in address D31 as a tare value)
get fixtare
X119=1
Save gross value as fixtare in address D31
set print
X120=1
Activate print order ?
clr marker 1
X64=0
Clear marker 1
clr marker 2
X65=0
Clear marker 2
clr marker 3
X66=0
Clear marker 3
select gross
X72=0
Save the gross weight in address D11
2. Determining a condition
Additionally, a condition [Condition] can be assigned to the signal; see Chapter 4.8.4.