X3 Operating Manual
Commissioning
Sartorius
97
4.9.1 Adapting the Analog Output
The output current can be adapted in small ranges. This is required, if small deviations from the nominal
value occur in a connected PLC.
Press and select [Show HW-slots]-[Slot 1/2 PR5510/07 analog I/O] to open the menu:
Info/HW-Slots
Builtin RS
232
Slot 1
PR 5510/07
analog I/O
PR 5510/07 is fitted in Slot 1
Slot 2
-empty-
Slot 3
Builtin
Digital I/O
Slot 4
PR 1721/31
Profibus-DP
Select [PR 5510/07 analog I/O] and confirm with
.
Info/HW-Slots
PR5510/07 on slot
1
In use by PLC task
2
Analog output
0.0000 mA
Select with
Stop PLC Stop I/O
Adjust
Reset
Press [Adjust]
Adapt analog output ?
Safety prompt: Reply [Yes] to start adapting.
Info/HW-Slots/Adjust Analog Output Slot 1
Output
4.000 mA
Measured
4.004 mA
Enter e.g. the value for 4 mA measured by the connected PLC under [Measured].
After pressing , the 2nd value (20 mA) is displayed:
Info/HW-Slots/Adjust Analog Output Slot 1
Output
20.000 mA
Measured
20.010 mA
Enter e.g. the value for 20 mA measured by the connected PLC under [Measured].
Save settings ?
Reply [Yes] to validate the changes, or
[No] to keep the previous values.
If you want to return to the factory settings (4 mA and 20 mA):
Press [Reset]: A safety prompt is displayed:
Reset to default ?
Reply [Yes] for reset to the factory settings, or
[No], if you want to keep the entered values unchanged.