Manual SA 235
page 11
Whether the 1 point or 3 point adjustment procedure is in operation can be
determined as follows:
1. switch the scale off if he is on
2. press together both the keys + and keep them pressed
3. switch the scale on with key and listen to the "beep"
4.1 if 1 x a "beep" sounds during switching on, the 1 point
procedure is in operation (H2 in the menu)
4.2 if 1.5 x a "beep" sounds during switching on, the 1 point
procedure is in operation (H1 in the menu, the 2
nd
beep is very short . .)
The selection of the adjustment procedure will be switched over any time
again this procedure is executed as per point 1 . . 3 (see above).
The adjustment mode may be entered as follows:
1. switch the scale off if he is on, switch off the count and check weighing.
2. press key and keep it pressed.
3. switch the scale on with key > 1 sec.
4. wait until de reading is zero, then release key
5. press together all the 3 keys + + until Sp appears.
6. now is shown the actual weighing capacity, it can be changed with key,
however regard the capacity of the loadcell ! Never select a higher
capacity, it will cause destruction of the loadcell !
7. confirm with key , het display toont: LS - - - -
8. start the zero point adjustment, using key again.
9. now the zero adjustment is executed, the most right display figure
runs down from 5 to 0 .
10. continue with the 1 point or the 3 points adjustment procedure, depending
on the display showing: H2 or: H1 .
1. the display shows H2 = 1 point adjustment procedure is selected.
2. now place accurate weights of 2/3 of the maximum weighing range
on the weigh platform. (e.g. for a 3 kg scale: place 2 kg)
3. confirm with key , automatically followed by:
4. the weight adjustment will be executed, the most right display figure
runs down from 5 to 0 .
5. the scale turns off automatically; remove the weight.
6. the scales switches itself on again in the weighing mode.
WEEGTECHNIEK HOLLAND B.V.
Website www.satex.nl
Tel +31 36 522 20 30
Patroonsweg 23
Email info@satex.nl
Fax +31 36 522 20 60
3892 DA ZEEWOLDE The Netherlands
ADJUSTING THE SCALE
Select: 1 point or 3 point adjustment
Entering the calibration / adjustment Mode
(in the lb mode to be executed in kg)
1 point adjustment: H2
returns
from 5 to 0
returns
from 5 to 0