background image

 

 

Precautions   

For safety and proper use of the bimetal thermometer, be sure to observe the following: 

 

- Properly insert the sensing stem into the object.    Otherwise, the correct measurement is not 

obtained.   

 
 
 
 
 

 
 

- In the following cases, use a well (option) for protecting sensing stem. 

    a. The liquid to be measured is more than 50°C. 

    b. Sensing stem gets pressure with more than 5 kg/cm2 

  (For the case diameter is 75 mm or less, use a well even if the pressure is 2 kg/cm2) 

    c. The object to be measured is corrosive.   

 

- Check regularly that the mounting screw and fixing nut are securely tightened. 

- Never disassemble or alter the thermometers, which may cause malfunctions. 

- Do not alter, bend or bind the sensing stem, as doing so may break or damage the 

thermometer. 

-To prevent injury, be careful of the sharp tip of the sensing stem especially when cleaning the 

stem after taking measurements. 

 

Maintenance 

For accurate measurements, it is desirable to confirm the accuracy of the bimetal 

thermometers in comparison with the standard thermometers once a year. 

 

 

WARRANTY POLICY 

Warranty 

Our products are warranted to be free from defects in materials and workmanship for a period of 

one year 

from date of delivery.    If repair or adjustment is necessary and has not been the result of abuse or misuse 

within the 

one-year-period

, please return the units on - Freight Prepaid – basis and correction of the defect 

will be made without charges.    We alone will determine if the product problem is due to deviations or 

customer misuse.   

                                      Out-of-warranty products will be repaired on charge basis.   

Return of items 

Authorization must be obtained from us before returning items for any reason. When applying for 

authorization, please include data regarding the reasons the items are to be returned.     

Please note that we reserve the right to make improvements in design, construction and appearance of our 

products without notice. 

 

 

 

3-4, Kanda-kajicho, Chiyoda-ku, Tokyo 101-0045 Japan 

URL: http://www.sksato.co.jp/english/ 

Co7

 

object measured 

 

Reviews: