Cautions
For your safety and proper use of the bimetal thermometer, be sure to observe the
following:
- Properly insert the sensing stem into object. Otherwise, the accuracy of measurement is
affected.
- In the following cases, it is desirable to use a well (option) for protecting sensing stem.
a. The liquid to be measured is more than 50°C.
b. More than 5 kg/cm2 pressure is applied on the sensing stem
* If the case dia. is less than Ø75 mm, use a well (option) when the pressure is
more than 2 kg/cm2.
c. The object to be measured is corrosive.
- Check regularly that the mounting screw and fixing nut are securely tightened.
- Never disassemble or alter the thermometer, which may cause malfunctions.
- Do not alter, bend or bind the sensing stem, as doing so may break or damage the
thermometers.
-To prevent injury, be careful of the sharp tip of the sensing stem especially when
cleaning the stem after taking measurements.
Maintenance
For accurate measurements, it is desirable to confirm the accuracy of the bimetal
thermometer in comparison with the standard thermometer once a year.