9
9. Average display function
The SK-1120 can display the average temperature (AVG). The unit displays the average
temperature measured (for up to approximately 9 hours) after the MAX/MIN key is pressed.
These values are obtained by summation averaging.
10. Resolution switching function
The resolution is automatically switched according to the temperature. For less than 190°C, the
resolution 0.1°C is used, and for 190°C or higher temperatures, the resolution 1°C is used.
CAUTIONS
- With the SK-1120, the decimal point is displayed when the indication resolution is 1°C.
11. Backlight function
The SK-1120 is provided with a backlight function. Pressing the key lights the LCD with a
green backlight. This permits the indication or other measurement data on the LCD to be read
easily. The backlight can be turned off by pressing the key again, or it is automatically
turned off one minute after pressing the key.
CAUTIONS
- Illuminating the backlight reduces the battery life.
- You may see a tiny spot on the polarizing plate of the backlight, but this is not a defect.
12. Temperature difference (T1 - T2) display function
Two sensor probes can be connected to connectors 1 and 2 of the SK1120 to measure
temperatures through two channels. The difference of temperatures (T1 - T2) obtained through
these channels can be displayed by pressing the sensor switch
key
. The relationships between
the upper and lower displays and the sensor switch key operations are shown below. While the
temperature difference, T1 - T2, is being displayed, the label "T1 - T2" is lit.
7 segments (Upper) Temp. (T1) displays
7 segments (Upper) Temp. (T2) displays
Power ON 7 segments (Lower) Temp. (T2) displays
7 segments (Lower) Temp. (T1) displays
7 segments (Upper) (T1-T2) displays
7 segments (Upper) (T1-T2) displays
7 segments (Lower) Temp. (T2) displays
7 segments (Lower) Temp. (T1) displays