10
CAUTION
Be sure to use the sensor probe with a type K thermocouple connector. Also, make sure the
sensor probes to be connected to connectors 1 and 2 on the main unit are of the same type
and have the same property.
13. Offset (REL) display function
With the SK-1120, the offset value relative to the temperature measured when the OFFSET
key (
) is pressed can be displayed (REL: relative value display function).
14. Time display function
With the SK-1120, the time elapsed since the TIME key was pressed is displayed on the time
display. The relationships between the time displays and the TIME key operations are shown
below:
If the MAX/MIN key is pressed after the TIME key was pressed, the MAX/MIN key has priority.
Therefore, the time elapsed since the latest maximum or minimum value was measured is
shown on the display.
15. Error Message
Display
Causes
Remedy
Probe is not connected
Connect the probe. K type thermocouple is
applicable
Incorrect thermocouple sensor
is connected
Connect the K type thermocouple sensor
Measured value is out of
display range
-50.1°C or more than 1301°C
Bring the value within the measuring range
Short circuit inside the probe
Stop using the instrument and replace with other
probe.
TIME key
TIME key
Measurement mode