11 

 

16. Troubleshooting 

 
Some of the following phenomena may not be considered defective. Check the phenomena 
carefully before requesting repair. 
 

Abnormal 

phenomenon 

Possible cause(s) 

Action(s) to be taken 

The power is not 
turned on 

The batteries have run out. 

Replace with new batteries. 

If the power will not turn on 
even after properly inserting 
new batteries, some part(s) 
within the main unit may have 
been damaged due to 
improper handling such as 
dropping or giving strong 
shock to the unit. 

If the parts have been damaged, ask 
your Sato sales representative or our 
Service Network for repair. 

The batteries are not properly 
installed. 

Make sure the orientation of the 
batteries is correct. Installing the 
batteries in the wrong direction is 
dangerous as the batteries might 
explode or liquid may leak. 

Indication value is 
abnormal. 

The batteries have run out. 

Replace with new batteries. 

The main unit is not 
accustomed to the ambient 
temperature. 

Leave the main unit in the ambient 
temperature for a while before starting 
the measurements. 

The sensor is being 
disconnected or the protection 
tube is deformed.   

Check the appearance of the sensor 
probe carefully. If any abnormality is 
found, ask us or our Service Network 

The sensor probe connector is 
not fully inserted. 

Fully insert the sensor probe connector. 

The sensor is not fully 
contacting the object to be 
measured. 

Review the sensor measurement 
procedures. If the object’ s thermal 
capacity is small, the indication may be 
affected by the protection tube size. 

The instrument is used in an 
environment where electrical 
noise is generated. 

Move the device to a place away from 
electrical noise. Installing shields may 
effectively block some electrical noises; 
wrapping the main unit and each of the 
sensor grip and cord with a sheet of 
aluminum foil may also help. 

The indication is 
unstable. 

The batteries have run out.   

Replace with new batteries 

The sensor is being 
disconnected or the protection 
tube is deformed. 

Check the appearance of the sensor 
probe carefully. If any abnormality is 
found, ask us or our Service Network   

Summary of Contents for Jumbo SK-1120

Page 1: ...Instruction Manual Jumbo LCD Digital Thermometer Model SK 1120 2 channel 3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 Japan URL http www sksato co jp english M 11 ...

Page 2: ......

Page 3: ...may result in unstable display or lower accuracy Never use the SK 1120 out of the measurement range as this may cause malfunctions Never wet or use the SK 1120 underwater They are not of waterproof construction When used outdoors keep the SK 1120 out of rain or water splashes If it becomes wet the unit may fail If the unit is not to be used for a long time always remove the batteries which may oth...

Page 4: ...ia these two probes are displayed on the upper display and the lower display and the difference between them T1 T2 can also be displayed on the upper display Average temperature display function The average value of the temperatures measured after pressing the AVG key can be displayed OFFSET key The offset value relative to the temperature when this key was pressed is displayed Backlight function ...

Page 5: ...rage mode Press the MAX MIN key in storage mode to switch the display of MAX and MIN Press the MAX MIN key in MAX and MIN display mode for 2 sec or more to return to normal measurement mode HOLD REL key Press the HOLD REL key in normal measurement mode to hold freeze temperature readings Press this key again to return to normal measurement mode Backlight switch key Turns on or off the backlight TI...

Page 6: ...re measured via sensor probe 1 T1 the temperature measured at sensor probe 2 T2 or the difference T1 T2 in sequence every time the sensor switch key is pressed 15 Seven segment display lower display Displays the temperature measured via the other sensor probe when the sensor switch key is pressed When the upper display shows T1 this display shows T2 or vice versa 16 C Indicates that the display un...

Page 7: ...st or lowest temperature and its occurrence time can be displayed together with the maximum or minimum or average value 25 Lights when temperature value is held 26 Low battery mark Lights when battery power is insufficient Replace the batteries with new ones 3 Storage Store the SK 1120 in an environment where the temperature range is within 10 C to 60 C and the relative humidity is not more than 8...

Page 8: ...line batteries 5 Fitting the vinyl cover over the unit The SK 1120 comes with a vinyl cover that fits on the unit With the cover in place the thermometer can be used even when your hands are wet The cover also keeps the unit clean and dry Cautions SK 1120 is not waterproof Do not place it in water or use it where it may be splashed with water Otherwise it will fail If water enters the vinyl cover ...

Page 9: ...in unit the LCD displays an OL or OL to warn that the sensor is overrange 3 You can avoid the influence of ambient temperature around the protection tube by inserting the sensor probe into the target regardless of whether gas liquid or solid to approximately 15 times the diameter of the protection tube from the end of the sensor protection tube approximately 35 mm for the SK K010 CAUTIONS Take car...

Page 10: ... the MAX MIN key for at least two seconds The time when the latest maximum or minimum value is recorded can be displayed in the time display area on the LCD Other uses Unless the normal measurement mode is restored the maximum or minimum value display mode continues keeping the latest maximum or minimum value on the display In this way the peak or valley hold function indication of the maximum or ...

Page 11: ...ff by pressing the key again or it is automatically turned off one minute after pressing the key CAUTIONS Illuminating the backlight reduces the battery life You may see a tiny spot on the polarizing plate of the backlight but this is not a defect 12 Temperature difference T1 T2 display function Two sensor probes can be connected to connectors 1 and 2 of the SK1120 to measure temperatures through ...

Page 12: ... display The relationships between the time displays and the TIME key operations are shown below If the MAX MIN key is pressed after the TIME key was pressed the MAX MIN key has priority Therefore the time elapsed since the latest maximum or minimum value was measured is shown on the display 15 Error Message Display Causes Remedy Probe is not connected Connect the probe K type thermocouple is appl...

Page 13: ...perature Leave the main unit in the ambient temperature for a while before starting the measurements The sensor is being disconnected or the protection tube is deformed Check the appearance of the sensor probe carefully If any abnormality is found ask us or our Service Network The sensor probe connector is not fully inserted Fully insert the sensor probe connector The sensor is not fully contactin...

Page 14: ...4 20 Number of channel 2 channel Functions Automatic resolution switching function MAX MIN maximum minimum memory data AVG Average value HOLD data hold OFFSET offset value Backlight Measurement of variation CH1 CH2 Auto Power OFF about 30 min Measuring range 50 to 1300 Accuracy main body 0 5 rdg 1 at 50 to 0 0 0 3 rdg 1 at 0 0 to 1000 0 5 rdg 1 at 1001 to 1300 Resolution 0 1 1 w resolution switchi...

Page 15: ...SK 1120 ...

Page 16: ......

Reviews: