12
The indication is
unstable (continued)
The sensor probe connector is
not fully inserted.
Fully insert the sensor probe connector.
The indication is
unstable (continued)
The sensor is not fully
contacting the object to be
measured.
Review the sensor measurement
procedures. If the object's thermal
capacity is small, the indication may be
affected by the protection tube size.
The instrument is used in an
environment where electrical
noise is generated.
Move the device to a place away from
electrical noise. Installing shields may
effectively block some electrical noises;
wrapping the main unit and each of the
sensor grip and cord with a sheet of
aluminum foil may also help.
17. Specifications
Name of Product
Digital Thermometer (K Type Thermocouple)
Model No.
SK-1120
Cat. No.
8014-20
Number of channel
2 channel
Functions
Automatic resolution switching function
MAX/ MIN: maximum/ minimum memory data
AVG: Average value
HOLD: data hold
OFFSET: offset value, Backlight
Measurement of variation (CH1 –
CH2)
Auto-Power OFF (about 30 min.)
Measuring range
-50 to 1300℃
Accuracy
(main body)
±(0.5%rdg + 1) ℃ at -50 to 0.0℃
±(0.3%rdg + 1) ℃ at 0.0 to 1000℃
±(0.5%rdg + 1) ℃ at 1001 to 1300℃
Resolution
0.1℃/1℃ (w/ resolution switching function)
Sampling time
About 0.5 sec.
Sensor
K Thermocouple, Connector: ANSI standard, Yellow
Accuracy (sensor)
K Thermocouple, class 2
Operation ambient
0 to 50℃ for temperature, less than 80%RH for humidity no condensing
Power requirements
6 AAA Alkaline batteries
Battery life
About 80 hours (Backlight OFF)
About 25 hours (Backlight ON)
* continuous with alkaline batteries
Materials
Main unit : ABS resin
Probe : PVC resin Sensor Cable: PV resin
Dimensions
Main unit : 72(W)
151(H)
35(D) mm
Weight
approx. 255g (including batteries)
Standard Accessories
“AAA”
Alkaline battery x 6, vinyl cover