 Methods for Measuring Temperature in Storage Mode 

- 1. On Setting Modes 

When you enter a setting mode, the numerical values that can be set 
start flashing. The numerical values can be changed by pressing the 
UP or DOWN key. To select other setting mode or quit the setting 
mode, press the SET / CLEAR key. The setting mode will switch back 
to normal measurement mode. 

-2. Setting Time 

Press the POWER and TIME keys simultaneously while the main unit 
is on. Set the time with the following procedure. After setting each 
item, press the SET / CLEAR key to move to set the next item.

Time setting is required for both manual and auto storage modes. 

Note :  If manual or auto storage mode is invoked without setting the 

current time, the factory-set date and time will be recorded in 
the stored data.

•  Example:When setting 10:53 am, April 1, 2016 

Year setting

Month setting

Day setting

is flashing

is flashing

is flashing

is flashing

is flashing

Setting complete

Minute setting

Hour setting

``   ´´

``     ´´

``     ´´

``     ´´

``           ´´

- 3. Setting the Recording Interval 

Press the POWER and HOLD / REL keys simultaneously while 
the main unit is on. Set the recording interval with the following 
procedure. 
Note : The recording interval needs to be set for auto storage mode. 

• The default recording interval is 1 minute.

If, for example, you wish to set the recording 
interval to 10 minutes, set the value as 
shown on the right by pressing the UP key. 
Press the SET / CLEAR key to switch back 
to normal measurement mode. If the current 
time is 13:06, the unit will start recording data from 13:10 because 
recording occurs every 10 minutes on the hour (see the following 
table). 

(min: minute, h: hour)

Recording interval

Recording start timing

1 min

Every minute on the hour

2 min
3 min

Every 10 minutes on the hour

5 min

10 min
30 min

1 h

Every hour on the hour

2 h
3 h
4 h
6 h

12 h
24 h

15

16

Summary of Contents for SK-1260

Page 1: ...ON MANUAL No 8080 05 WATERPROOF DIGITAL THERMOMETER With Print Output MODEL SK 1260 3 4 Kanda Kajicho Chiyodaku Tokyo 101 0045 JAPAN TEL 81 0 3 3254 8117 FAX 0 3 3254 8123 URL http www sksato co jp english Q 11 ...

Page 2: ...ses Read this manual thoroughly before using Keep the manual in a safe place for future references whenever necessary Overview The SK 1260 is a waterproof digital thermometer to which dedicated thermistor sensor probe and dedicated thermocouple K sensor probe are connected Measured data can be stored in the main unit and can be printed on an optional printer ...

Page 3: ...Error Message 22 V Specifications 23 VI Warranty Policy 24 Warning The SK 1260 is not explosion proof Never use it in an atmosphere containing flammable gases Beware of explosion There is a risk of explosion Take extreme care If you have any questions contact the dealer from which the unit was purchased or our service network Features Waterproof Function The main unit connected with probe has a wa...

Page 4: ... and MAX MIN keys simultaneously in normal measurement mode to start auto storage mode While the main unit is on press the REC PRINT and POWER keys simultaneously to print the stored data on an optional printer LCD panel The LCD panel displays the temperature reading the current state of settings and characters Sensor connector Connect the dedicated thermistor sensor probe or dedicated thermocoupl...

Page 5: ...r errors Do not forcibly pull bend or bundle the sensor cord Placing a heavy object on the sensor cord heating or scratching the cord may also damage it If this unit is not to be used for a long time always remove the batteries from the unit Otherwise the batteries may leak fluid resulting in failure p o ⓐ Battery mark Lit when battery power is insufficient ⓑ TIME Lit while time is being set ⓒ Th ...

Page 6: ... sounds for one second every one minute Do not wash or wipe this unit with alcohol thinner or other solvents Also do not wash it in water If the unit becomes dirty wipe it with a tightly wrung cloth that has been dipped in warm water with neutral detergent The sensors are specially designed for the SK 1260 Water proof Digital Thermometer Never connect these to another thermometer or a personal com...

Page 7: ...re to turn the power off before replacing the batteries Otherwise all data will be cleared 5 Connecting the Sensor Probe Insert the plug of the sensor probe all the way into the connector socket on the main unit until clicks are heard Note Hold the sensor plug with the SK mark facing up so that it becomes level with the front panel then insert the plug all the way in Failure to do so affects the w...

Page 8: ...ter on then press the POWER and REC keys simultaneously Printing will start Temperature measurement in manual or auto storage mode Displaying stored data on the LCD panel Normal measurement mode Time setting mode Select storage mode Manual storage mode Start storing data Auto storage mode Set the recording interval Set management numbers Start storing data Call stored data Normal measurement mode ...

Page 9: ...lue If the temperature changes too frequently during measurement press the HOLD REL key to hold freeze the measurement data on the main 7 segment display for easier reading The currently measured temperature is continuously displayed on the sub 7 segment display so that you can check the current temperature while holding the necessary temperature reading Press the HOLD REL key again to invoke REL ...

Page 10: ...he stored values have been cleared Note While the maximum or minimum temperature is being displayed the current temperature is displayed on the sub 7 segment display Ⅰ 6 Auto Power off Function The power will be turned off automatically if the unit is not operated for approximately 30 minutes This function conserves battery power if you forget to turn the power off The following table shows whethe...

Page 11: ... 10 53 am April 1 2016 Year setting Month setting Day setting is flashing is flashing is flashing is flashing is flashing Setting complete Minute setting Hour setting Ⅱ 3 Setting the Recording Interval Press the POWER and HOLD REL keys simultaneously while the main unit is on Set the recording interval with the following procedure Note The recording interval needs to be set for auto storage mode T...

Page 12: ...e TIME key for at least one second to display the stored temperature values on the main 7 segment display and the file numbers on the sub 7 segment display The temperature values and file numbers are displayed in chronological order starting with the first file Hold down the TIME key to advance the temperature values and file numbers rapidly Press the SET CLEAR key to switch back to normal measure...

Page 13: ...tart timing Display interval Display time 1 min 2 min 3 min Every min on the hour Continuous display 5 min 10 min Every 10 min on the hour 30 min Every 30 sec For about 7 sec 1 hour Every hour on the hour Every 1 min 2 hours Every 2 min 3 hours Every 3 min 4 hours Every 5 min 6 12 and 24 hours Every 10 min Be sure to turn the power off by pressing the POWER key in following procedures Switching th...

Page 14: ...LCD Display Causes Remedy Er Hi Measured value is higher than measuring range Stop measurement and bring the value within the measuring range Er Lo Measured value is lower than measuring range Measured value exceeds the measuring range further appears after Hi or Lo displayed Probe is not connected Connect the probe Sensor is disconnected Poor connection of probe connector or disconnection inside ...

Page 15: ...t notice Ⅴ Specifications Product No 8080 05 Waterproof Digital Thermometer Model SK 1260 Measuring range Thermistor 30 0 to 199 9 C K Thermocouple 99 9 to 1250 C Resolution Thermistor 0 1 C at 30 0 to 199 9 C K Thermocouple 0 1 C at 99 9 to 199 9 C 1 C at 200 C to 1250 C Accuracy Thermistor Main unit 0 2 C Sensor 0 1 rdg 0 2 C at 0 0 to 150 0 C 0 2 rdg 0 4 C at 30 0 to 0 1 C 150 1 to 199 9 C K Th...

Page 16: ......

Reviews: