Ⅲ
-3. Examples of the Printed Data
Manual storage mode
Auto storage mode (Recording interval is set to 2 min.)
Rec
Date
Time
Temp
Print-Date : 2016 / 9 / 23
Mes-Code : No. 00-00 (°C)
Measuring Data
Rec
Date
Time
Ave
Max
Min
Print-Date : 2016 / 9 / 23
Sensor : K1, Mes-Code : No. 00-00 (°C)
Measuring Data
Ⅳ
Error Message
• Error massage is displayed on the main 7-segment LCD
Display
Causes
Remedy
Er : Hi
Measured value is higher
than measuring range.
Stop measurement and
bring the value within the
measuring range.
Er : Lo
Measured value is lower
than measuring range.
Measured value exceeds
t h e m e a s u r i n g r a n g e
further.
(appears after Hi
or Lo displayed)
Probe is not connected.
Connect the probe.
Sensor is disconnected.
Poor connection of probe
connector or disconnection
inside the probe.
If disconnection seems to
be the cause, stop using the
thermometer.
Short-circuits inside the
probe.
Stop using the thermometer.