Ⅰ
General Temperature Measurement Methods
Ⅰ
-1. Temperature Measurement Method
⑴
Connect the sensor probe to the main unit.
⑵
Press the POWER key for at least one second. Test patterns will
appear on the entire LCD panel for approximately five seconds,
then the currently measured temperature will be displayed in
normal measurement mode.
*
If the unit is not operated for 30 minutes, the Auto power-off
function will be activated, turning the power off automatically.
If the POWER key is pressed for 3 seconds or more when the
unit is turned on, Auto power-off
function is released and the unit
enters normal measurement. In this case,
mark is displayed
to show that the Auto power-off function is not being set.
Caution!
・…
When measuring high temperature, be careful not to burn yourself.
・
Do not disconnect the sensor probe while the power is on.
⑶
When an accurate temperature measurement is needed, insert the
sensor into the measuring object to a depth of approximately 15
times the diameter of the sensor stem, regardless of whether the
object is a gas, liquid, or semi-solid. For example, if the sensor
stem is 2 mm in diameter, insert the sensor into the object to a
depth of approximately 30 mm or more from the tip of the stem.
This will shield the sensor from the ambient temperature.
⑷
Press the POWER key for at least one second in normal
measurement mode to turn off the display and main unit.
Ⅰ
-2. HOLD / REL Function
⑴
Press the HOLD / REL key to hold (freeze) the measured value.
If the temperature changes too frequently during measurement,
press the HOLD / REL key to hold (freeze) the measurement data
on the main 7-segment display for easier reading. The currently
measured temperature is continuously displayed on the sub
7-segment display so that you can check the current temperature
while holding the necessary temperature reading.
⑵
Press the HOLD / REL key again to invoke REL mode in which the
fluctuation between the currently measured value and the held
value appears on the main 7-segment display. Temperatures lower
than - 199.9°C are displayed as "OL". Measured values continue
to appear on the sub 7-segment display in real-time. Each time
measured values change, the difference between the currently
measured value and the held value appears on the main 7-segment
display.
⑶
Press the HOLD / REL key again to return to normal measurement
mode.
*
Release HOLD and Current temperature (sub 7-segment)
display functions
By pressing the SET / CLEAR and the HOLD / REL keys
simultaneously in the HOLD function, the HOLD mark and the
current temperature displayed on sub 7-segment will disappear.
Press HOLD / REL key to appear REL display and press
HOLD / REL key again to return to normal measurement mode.
Ⅰ
-3. MAX / MIN Function
• Displaying the maximum and minimum temperatures
⑴
Press the MAX / MIN key once in normal measurement mode, then
press the SET / CLEAR key to start MAX / MIN function.
⑵
Press the MAX / MIN key to display the lowest temperature since
the MAX / MIN function started.
⑶
Press the MAX / MIN key to switch back in normal measurement
mode then press the MAX / MIN key again to display the highest
temperature since the MAX / MIN function started.
11
12
Summary of Contents for SK-1260
Page 16: ......