background image

  General Temperature Measurement Methods

-1. Temperature Measurement Method

 

Connect the sensor probe to the main unit. 

Press the POWER key for at least one second. Test patterns will 
appear on the entire LCD panel for approximately five seconds, 
then the currently measured temperature will be displayed in 
normal measurement mode.

*

If the unit is not operated for 30 minutes, the Auto power-off 

function will be activated, turning the power off automatically. 
If the POWER key is pressed for 3 seconds or more when the 
unit is turned on, Auto power-off

 

function is released and the unit 

enters normal measurement. In this case, 

mark is displayed 

to show that the Auto power-off function is not being set.

Caution!

・…

When measuring high temperature, be careful not to burn yourself. 

Do not disconnect the sensor probe while the power is on.

When an accurate temperature measurement is needed, insert the 

sensor into the measuring object to a depth of approximately 15 
times the diameter of the sensor stem, regardless of whether the 
object is a gas, liquid, or semi-solid. For example, if the sensor 
stem is 2 mm in diameter, insert the sensor into the object to a 
depth of approximately 30 mm or more from the tip of the stem. 
This will shield the sensor from the ambient temperature. 

Press the POWER key for at least one second in normal 

measurement mode to turn off the display and main unit.

-2. HOLD / REL Function 

Press the HOLD / REL key to hold (freeze) the measured value.
If the temperature changes too frequently during measurement, 
press the HOLD / REL key to hold (freeze) the measurement data 
on the main 7-segment display for easier reading. The currently 

measured temperature is continuously displayed on the sub 
7-segment display so that you can check the current temperature 
while holding the necessary temperature reading. 

Press the HOLD / REL key again to invoke REL mode in which the 

fluctuation between the currently measured value and the held 
value appears on the main 7-segment display. Temperatures lower 
than - 199.9°C are displayed as "OL". Measured values continue 
to appear on the sub 7-segment display in real-time. Each time 
measured values change, the difference between the currently 
measured value and the held value appears on the main 7-segment 
display. 

Press the HOLD / REL key again to return to normal measurement 

mode.

*

 Release HOLD and Current temperature (sub 7-segment) 
display functions

By pressing the SET / CLEAR and the HOLD / REL keys 
simultaneously in the HOLD function, the HOLD mark and the 
current temperature displayed on sub 7-segment will disappear. 
Press HOLD / REL key to appear REL display and press 
HOLD / REL key again to return to normal measurement mode. 

-3. MAX / MIN Function 

• Displaying the maximum and minimum temperatures 

Press the MAX / MIN key once in normal measurement mode, then 

press the SET / CLEAR key to start MAX / MIN function. 

Press the MAX / MIN key to display the lowest temperature since 

the MAX / MIN function started. 

Press the MAX / MIN key to switch back in normal measurement 
mode then press the MAX / MIN key again to display the highest 
temperature since the MAX / MIN function started.

11

12

Summary of Contents for SK-1260

Page 1: ...ON MANUAL No 8080 05 WATERPROOF DIGITAL THERMOMETER With Print Output MODEL SK 1260 3 4 Kanda Kajicho Chiyodaku Tokyo 101 0045 JAPAN TEL 81 0 3 3254 8117 FAX 0 3 3254 8123 URL http www sksato co jp english Q 11 ...

Page 2: ...ses Read this manual thoroughly before using Keep the manual in a safe place for future references whenever necessary Overview The SK 1260 is a waterproof digital thermometer to which dedicated thermistor sensor probe and dedicated thermocouple K sensor probe are connected Measured data can be stored in the main unit and can be printed on an optional printer ...

Page 3: ...Error Message 22 V Specifications 23 VI Warranty Policy 24 Warning The SK 1260 is not explosion proof Never use it in an atmosphere containing flammable gases Beware of explosion There is a risk of explosion Take extreme care If you have any questions contact the dealer from which the unit was purchased or our service network Features Waterproof Function The main unit connected with probe has a wa...

Page 4: ... and MAX MIN keys simultaneously in normal measurement mode to start auto storage mode While the main unit is on press the REC PRINT and POWER keys simultaneously to print the stored data on an optional printer LCD panel The LCD panel displays the temperature reading the current state of settings and characters Sensor connector Connect the dedicated thermistor sensor probe or dedicated thermocoupl...

Page 5: ...r errors Do not forcibly pull bend or bundle the sensor cord Placing a heavy object on the sensor cord heating or scratching the cord may also damage it If this unit is not to be used for a long time always remove the batteries from the unit Otherwise the batteries may leak fluid resulting in failure p o ⓐ Battery mark Lit when battery power is insufficient ⓑ TIME Lit while time is being set ⓒ Th ...

Page 6: ... sounds for one second every one minute Do not wash or wipe this unit with alcohol thinner or other solvents Also do not wash it in water If the unit becomes dirty wipe it with a tightly wrung cloth that has been dipped in warm water with neutral detergent The sensors are specially designed for the SK 1260 Water proof Digital Thermometer Never connect these to another thermometer or a personal com...

Page 7: ...re to turn the power off before replacing the batteries Otherwise all data will be cleared 5 Connecting the Sensor Probe Insert the plug of the sensor probe all the way into the connector socket on the main unit until clicks are heard Note Hold the sensor plug with the SK mark facing up so that it becomes level with the front panel then insert the plug all the way in Failure to do so affects the w...

Page 8: ...ter on then press the POWER and REC keys simultaneously Printing will start Temperature measurement in manual or auto storage mode Displaying stored data on the LCD panel Normal measurement mode Time setting mode Select storage mode Manual storage mode Start storing data Auto storage mode Set the recording interval Set management numbers Start storing data Call stored data Normal measurement mode ...

Page 9: ...lue If the temperature changes too frequently during measurement press the HOLD REL key to hold freeze the measurement data on the main 7 segment display for easier reading The currently measured temperature is continuously displayed on the sub 7 segment display so that you can check the current temperature while holding the necessary temperature reading Press the HOLD REL key again to invoke REL ...

Page 10: ...he stored values have been cleared Note While the maximum or minimum temperature is being displayed the current temperature is displayed on the sub 7 segment display Ⅰ 6 Auto Power off Function The power will be turned off automatically if the unit is not operated for approximately 30 minutes This function conserves battery power if you forget to turn the power off The following table shows whethe...

Page 11: ... 10 53 am April 1 2016 Year setting Month setting Day setting is flashing is flashing is flashing is flashing is flashing Setting complete Minute setting Hour setting Ⅱ 3 Setting the Recording Interval Press the POWER and HOLD REL keys simultaneously while the main unit is on Set the recording interval with the following procedure Note The recording interval needs to be set for auto storage mode T...

Page 12: ...e TIME key for at least one second to display the stored temperature values on the main 7 segment display and the file numbers on the sub 7 segment display The temperature values and file numbers are displayed in chronological order starting with the first file Hold down the TIME key to advance the temperature values and file numbers rapidly Press the SET CLEAR key to switch back to normal measure...

Page 13: ...tart timing Display interval Display time 1 min 2 min 3 min Every min on the hour Continuous display 5 min 10 min Every 10 min on the hour 30 min Every 30 sec For about 7 sec 1 hour Every hour on the hour Every 1 min 2 hours Every 2 min 3 hours Every 3 min 4 hours Every 5 min 6 12 and 24 hours Every 10 min Be sure to turn the power off by pressing the POWER key in following procedures Switching th...

Page 14: ...LCD Display Causes Remedy Er Hi Measured value is higher than measuring range Stop measurement and bring the value within the measuring range Er Lo Measured value is lower than measuring range Measured value exceeds the measuring range further appears after Hi or Lo displayed Probe is not connected Connect the probe Sensor is disconnected Poor connection of probe connector or disconnection inside ...

Page 15: ...t notice Ⅴ Specifications Product No 8080 05 Waterproof Digital Thermometer Model SK 1260 Measuring range Thermistor 30 0 to 199 9 C K Thermocouple 99 9 to 1250 C Resolution Thermistor 0 1 C at 30 0 to 199 9 C K Thermocouple 0 1 C at 99 9 to 199 9 C 1 C at 200 C to 1250 C Accuracy Thermistor Main unit 0 2 C Sensor 0 1 rdg 0 2 C at 0 0 to 150 0 C 0 2 rdg 0 4 C at 30 0 to 0 1 C 150 1 to 199 9 C K Th...