−59 −
(5) Press the REC/STOP key to confirm. Next, set the lower limit for
temperature.
Tip: If the lower limit is not necessary, press the CALL/OUT key once.
The SK-L200T II or SK-L200TH II
α
returns to the measurement state
for temperature or temperature and humidity, respectively.
(6) The Alarm and Hi marks light. Press the
▲
and
▼
keys at the same time.
Note: Pressing the
▲
key and
▼
key simultaneously one more time will turn
the warning function off.
(7) Use the
▲
or
▼
key to display the desired value for lower limit.
(8) Press the REC/STOP key to confirm.
Tips: If the lower limit is not necessary, press the CALL/OUT key once.
The SK-L200T II or SK-L200TH II
α
returns to the measurement state
for temperature or temperature and humidity, respectively. For setting
the humidity upper/lower limit in the SK-L200TH II
α
model, repeat
steps (3) to (8) above.
(9) Press the CALL/OUT key once to complete the setting.
The SK-L200T II or SK-L200TH II
α
returns to the measurement state for
temperature or temperature and humidity, respectively.