−72 −
Action
・
Check if the probe is properly
connected to the Datalogger.
・
Check if the probe is properly
connected to the Datalogger.
・
Check if "Hi", "Lo" or "Er" is
displayed on the upper LCD.
・
Humidity reading may be out of
the measuring range.
・
Replace the batteries with new
ones.
・
Erase the data stored on the
Datalogger.
If necessary, store the data
before erasing.
Description
Cannot measure temperature.
Cannot measure humidity.
Battery power problems:
・
The preset start was set when
the battery power was low.
・
The battery power became low
while in standby for preset start.
・
The battery power became low
while logging.
・
The logging option chosen was
different from that used in the
data currently stored in the
Datalogger.
Error code
E r
(upper LCD)
E r
(lower LCD)
(only
SK-L200 TH II
α
)
E r 1
E r 3
8. Troubleshooting
If an error occurs, one of the error codes listed below will be displayed on the
LCD. Check the details of the error and take corrective actions as appropriate.
If the problem persists, please contact the store where you purchased this
product or our Service Network.