Planning, transport, installation
LASERTEC 40
Page 50
Prepare cooling plant for transportation
As far as the cooling plant consists of a multitude of sensitive electronic
and mechanical components, observe the following preconditions and
instructions for storage, installation and transportation of the plant.
Transportation
The transportation of the cooling plan must be performed as careful and
shock-free and vibration-free as possible. Please observe the following
measures:
•
The cooling plant must be completely emptied before transportation.
•
Use only suitable packing material - shock-resistant and highly
vibration-absorbent to ensure a safe transportation of the sensitive
components.
•
Furthermore, the packing must provide a protection of the cooling
plant from dust and soiling.
•
Pack the cooling plant in such manner to prevent it from shocks and
falling.
•
If the cooling plant is shipped separately, use the original packing
bolster.
•
Required marking of the transported goods includes ”Protect against
humidity”, ”Transport and store in vertical position” and ”Fragile”.
Danger of freezing!
Before the transportation the cooling plant must be completely emptied,
otherwise danger of damage by freezing cooling water.
Storage
•
Always empty the cooling plant before storage.
•
The cooling plant must be stored horizontally only.
•
Ensure protection against dust and moisture.
•
For storage, the following ambient conditions must be observed:
Temperature:
5 °C to + 65 °C
Relative air humidity:
10 – 80 %, Not condensing