3 Technical features
3.1 Technical features of the devices
KT 320
KTT 320
Units displayed
°C, °F, °Ctd, °Ftd, %RH, mV, V, mA, A
Programmed and free units are also
available
1
(see table page 9)
°C, °F
Resolution
0.1°C, 0.1°F, 0.1%RH, 1 mV, 0.001 V,
0.001 mA, 0.1 A
0.1°C, 0.1°F
External input
Female micro-USB connector
Input for probe
2 SMART PLUG
2
inputs
4 inputs for thermocouple probes
(K, J, T, N, S)
Internal sensor
Temperature
-
Type of sensor
CTN
Thermocouple
Measuring range
Measuring range of the internal sensor
3
:
From -40 to +70°C
K: from -200 to +1300°C
J: from -100 to +750°C
T: from -200 to +400°C
N: from -200 to +1300°C
S: from 0 to 1760°C
Accuracy
4
±0.4°C from -20 to 70°C
±0.8°C below -20°C
K, J, T, N: ±0.4°C from 0 to 1300°C
±(0.3% of the r0.4°C) below 0°C
S: ±0.6°C
Setpoints alarm
2 setpoint alarms on each channel
Frequency of
measurements
From 1 second to 24 hours
Operating temperature
From -40 to +70°C
From -20 to 70°C
Storage temperature
From -40 to +85°C
Battery life
7 years
5
European directives
2011/65EU RoHS II; 2012/19/EU WEEE; 2014/30/EU EMC; 2014/35/EU
1
Some units are available only with optional probes.
2
Input which allows to plug different SMART PLUG compatible probes: see optional probes and cables page 10.
3
Other measuring ranges are available according to the connected probe: see optional probes and cables page 10.
4
All accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurement carried out in the same conditions, or carried out with
calibration
compensation.
5
On the basis of 1 measurement each 15 minutes at 25°C.
Technical features
7