8
2.2.3 Omega Brackets Installation
Omega brackets are used to ensure the required distance from the wall is maintained.
1.
Attach the bracket (1) to its place – sync the
hole in the bracket(1) to the hole in the unit
2.
Screw the M6 bolts into the designated
places (the holes on top and below the
bracket(1))
3.
Repeat steps 1 and 2 with the other
bracket(1)
1
2.2.4 Remote Alarm Contacts Connection
External alarm can be connected to the remote
alarm terminal located at the back of the freezer
base above the power cord. The alarm is outputted
from this terminal with contact capacity of 30 VDC,
2 A. The contact output is as follows:
Between
COM and N.O.
Between
COM and N.C
At normal
Open
Close
At abnormal
Close
Open
Note:
The alarm is freezer temperature-driven and only
actuated when the freezer temperature exceeds the high
or low temperature alarm limit.
2.2.5 Electrical Supply Connection and
Power Protection Bracket Installation
Plug the power cord into the power inlet at the rear
of the unit. Turn on the power switch. Install the
power protection bracket over the power cord and
power switch by installing the 3 screws into the pre-
tapped holes.
COM
N.O.
N.C.
(REMOTE ALARM TERMINAL)