1. Products Covered
Scancool Classic Ultra-Low Temperature Freezer
Model
Storage Volume
Temperature Range
Freezer Control
Electrical Rating
SCL510
439.7 Liter / 15.5 ft
3
-50
o
C to -86
o
C
Advanced Digital
Controller
230 VAC, 50 Hz, 1 Φ
SCL610
552.8 Liter / 19.5 ft
3
SCL710
668.8 Liter / 23.6 ft
3
SCL810
793.6 Liter / 28.0 ft
3
2. Safety Warning
Anyone working with, on or around this equipment should read this manual. Failure to read,
understand and follow the instructions given in this documentation may result in damage to the unit,
injury to operating personnel, and / or poor equipment performance.
Any internal adjustment, modification or maintenance to this equipment must be undertaken by
qualified service personnel.
The use of any hazardous materials in this equipment must be monitored by an industrial hygienist,
safety officer or some other suitably qualified individual.
Before you proceed, you should thoroughly understand the installation procedures and take note of
the environmental / electrical requirements.
In this manual, important safety related points will be marked with the symbol
If the equipment is used in a manner not specified by this manual, the protection provided by this
equipment may be impaired.
3. Important User Information
Battery switch located behind the right bottom door should always be turned "ON" when the unit is
running to charge the battery.
Low battery alarm may occur in the first 24 hours of operation until battery is charged.
CAUTION! To have the highest level of protection on stored samples, the use of a completely
independent backup system and a 24-hour redundant monitoring system with alarm capability is
recommended.
If power supply is intentionally cut off / disrupted, turn "OFF" the battery switch to protect the
battery from discharge damage.
4. Limitation of Liability
The disposal and / or emission of substances used in connection with this equipment may be governed by
various local regulations. Familiarization and compliance with any such regulations are the sole responsibility
of the users. Factory’s liability is limited with respect to user compliance with such regulations.