SHC 2000
9.900.900.719
SHC2000-manual-111609 11
3.2. Bath fluids
Caution:
Carefully read the safety data sheet of the bath fluid used, particularly
with regard to the fire point!
If a bath fluid with a fire point of
≤
65 °C is used, only supervised
operation is possible.
Water:
The quality of water depends on local conditions.
Ferrous water can cause corrosion - even on stainless steel.
Chloric water can cause pitting corrosion.
No liability for use with water.
Danger of freezing at working temperatures <5 °C.
Recommended bath fluids:
Bath fluids
Temperature range
Water
5 °C ... 80 °C
Mixture water/glycol
-30 °C ... 50 °C
Silicon oil
-40 °C to 150 °C
The maximum permissible viscosity is 30 mm
2
/s
Notice:
Please contact ScanCool before using other than recommended bath
fluids. ScanCool takes no responsibility for damages caused by the
selection of an unsuitable bath fluid.
Unsuitable bath fluids are liquids which e.g.
•
are very highly viscous
(much higher than 30 mm
2
/s [30 cSt] at the respective working
temperature)
•
have corrosive characteristics or
•
tend to cracking.
Caution:
The temperature controlling i.e. of fluids in a reactor constitutes normal
circulator practice.
We do not know which substances are contained within these vessels.
Many substances are:
•
inflammable,
easily
ignited or explosive
•
hazardous to health
•
environmentally
unsafe
i.e.:
dangerous
The user alone is responsible for the handling of these substances!