SHC 2000
20
•
The excess temperature value lies below the working
temperature setpoint.
Set the excess temperature to a higher value.
•
Cable of the excess temperature sensor interrupted or short-
circuited.
After eliminating the malfunction, press the mains power switch off
and on again to cancel the alarm state.
If the unit cannot be returned to operation, contact an authorized
service station.
Disturbances that are not indicated.
The electronic pump motor is overload-protected by an electronic
current limiter. If viscosity of the bath fluid is or becomes too high,
the motor stops running.
15
A
M
P
Mains circuit breakers (resettable) 15 A.