6 Troubleshooting
Error/display
Cause
Remedy
Flow detected during Warm-Up
Breath sample has been blown too early.
Deliver breath sample when requested by the instru-
ment.
Blow stronger!
/
Repeat test
Breath sample has been blown too soft.
Repeat test blowing harder.
Blow softer!
/
Repeat test
Breath sample has been blown too hard.
Repeat test blowing softer.
Blow longer!
/
Repeat test
Breath sample has been blown too short.
Repeat test blowing longer.
Suck after blow
/
Repeat test
Breath sample is not delivered correctly by blowing
and sucking.
Repeat test with blowing and immediately afterwards
sucking.
Suck longer!
/
Repeat test
Breath sample has been sucked too short.
Repeat test sucking longer.
Check mouthpiece
/
Repeat test
Mouthpiece is not correctly inserted into handset.
Insert mouthpiece correctly.
Repeat test
During delivery of the breath sample a fault has
been detected.
Deliver breath sample again when requested by the
instrument.
Test failed
The measured breath alcohol concentration is
above the set limit value.
Repeat the test after at least 15 minutes. If the mes-
sage appears, the breath alcohol concentration is
above the limit value.
Low temperature
/
Warm up handset
Temperature of the handset too low.
Detach handset and take it to a place with a moderate
temperature. Connect handset again and perform
test.
High temperature
/
Cool down handset
Temperature of handset too high.
Detach handset and take it to a place with a moderate
temperature. Connect handset again and perform
test.
Service required
in
X
days
Service interval will expire in x days.
Call an authorized Dräger Interlock service centre in
due time.
End of service
Service interval expired.
Call an authorized Dräger Interlock service centre.
Number of abnormalities
Too many violations have been detected.
Call an authorized Dräger Interlock service centre in
due time.