Mode : TD (Time Distance)
Triggered on: Flank
Waveform: Full (rectified)
Unit: Inches
Transducer: Single element
Frequency: 5.0 MHz
Gain: 37.0
Gain Increment:1 (meaning when you press the + key the gain will increase by 1
Mode
•
TD (Time Distance) is typical for time based measurements.
•
E
-
>E (Echo to Echo) The AScope is smart enough that when you turn on Gate2 it assumes you are
going to do Echo to Echo and will change modes for you. If you turn G2 off it will assume that you
no longer want Echo to Echo and will go back to TD setting.
•
FD (Flaw detect)
Trig
(Triggered on)
•
Flank (recommended for thickness measurements) Some peaks are out of range and therefore
distort your time based measurements
•
Peak
•
Z/C (Zero crossing)
Unit
(display units)
•
In (Inches)
•
mm (millimeters)
•
Us (microseconds)
W/F
(Waveform)
•
Full (rectified) displays the absolute value of the waveform data
•
RF (radio frequency) displays the signed value of the waveform data
•
+V Displays only the positive half wave
•
-
V Displays only the negative half wave
Trans
(Tranducer)
•
Single
Push drop down arrow to see
options of UT
UT Setup Settings